Park, Choi, Lee, and Kim: Central Corneal Thickness Measured by Four Different Methods in Normal and Post-Femtosecond Laser-Assisted LASIK Eyes

Abstract

Purpose

To compare corneal pachymetry assessment using four measurement methods in normal and post-femtosecond laser-assisted LASIK eyes.

Methods

Central corneal thickness was measured sequentially using Orbscan II, Pentacam, Galilei and ultrasonic pachymetry in 30 normal, non-surgical eyes (Group I), 30 eyes one to six months after femtosecond laser-assisted LASIK (Group II), and 30 eyes six months or longer after femtosecond laser-assisted LASIK (Group III).

Results

In Group I, corneal thickness measurements were similar for all four methods (P=0.202, one way ANOVA). In Groups II and III, corneal thickness measurements were significantly different (P=0.000, respectively, one way ANOVA). Compared to the Pentacam, Galilei and ultrasonic pachymetry, Orbscan significantly underestimated the corneal thicknesses in Groups II and III (P<0.005, respectively, one way ANOVA).

Conclusions

Central corneal thicknesses of normal eyes were similar for all four measurements, therefore corneal thickness measurements before refractive surgery using all four measurements is suitable. However measurements obtained with the Orbscan II were thinner than those obtained with the Pentacam, Galilei or ultrasonic pachymetry in post femtosecond laser-assisted LASIK eyes. Further studies are needed to determine which instrument is more accurate in measuring central corneal thickness before and after refractive surgery.

References

1. Binder PS. Ectasia after laser in situ keratomileusis. J Cataract Refract Surg. 2003; 29:2419–29.
crossref
2. Price FW Jr, Koller DL, Price MO. Central corneal pachymetry in patients undergoing laser in situ keratomileusis. Ophthalmology. 1999; 106:2216–20.
3. Iskander NG, Peters NT, Penno EA, Gimbel HV. Postoperative complications in laser in situ keratomileusis. Curr Opin Ophthalmol. 2000; 11:273–9.
crossref
4. Miglior S, Albe E, Guareschi M, et al. Intraobserver and interobserver reproducibility in the evaluation of ultrasonic pachymetry measurements of central corneal thickness. Br J Ophthalmol. 2004; 88:174–7.
crossref
5. Solomon OD. Corneal indentation during ultrasonic pachometry. Cornea. 1999; 18:214–5.
crossref
6. Rufer F, Schroder A, Arvani MK, Erb C. Central and peripheral corneal pachymetry–standard evaluation with the Pentacam system. Klin Monatsbl Augenheilkd. 2005; 222:117–22.
7. Menassa N, Kaufmann C, Goggin M, et al. Comparison and reproducibility of corneal thickness and curvature readings obtained by the Galilei and the Orbscan II analysis systems. J Cataract Refract Surg. 2008; 34:1742–47.
crossref
8. Kang PS, Kim JD, Yang YS. Comparison of corneal thickness measurements with the orbscan and ultrasonic pachymetry. J Korean Ophthalmol Soc. 2000; 41:1697–703.
9. Kim SH, Cho JH, Song BJ. Accuracy of Orbscan Pachymetry Measurements and Ultrasonic Pachymetry before and after LASIK with Orbscan IIⓇ Topography. J Korean Ophthalmol Soc. 2002; 43:2513–8.
10. Kwon M, Seoung Y, Hur D. Pachymetric measurements using orbscan after excimer refractive surgery. J Korean Ophthalmol Soc. 2004; 45:899–907.
11. Shin YJ, Kim NH, Kim DH. Comparison of Pentacam with Orbscan. J Korean Ophthalmol Soc. 2007; 48:637–41.
12. Ho T, Cheng AC, Rao SK, et al. Central corneal thickness measurements using Orbscan II, Visante, ultrasound, and Pentacam pachymetry after laser in situ keratomileusis for myopia. J Cataract Refract Surg. 2007; 33:1177–82.
crossref
13. Hashemi H, Mehravaran S. Central corneal thickness measurement with Pentacam, Orbscan II, and ultrasound devices before and after laser refractive surgery for myopia. J Cataract Refract Surg. 2007; 33:1701–7.
crossref
14. Kim SW, Byun YJ, Kim EK, Kim TI. Central corneal thickness measurements in unoperated eyes and eyes after PRK for myopia using Pentacam, Orbscan II, and ultrasonic pachymetry. J Refract Surg. 2007; 23:888–94.
crossref
15. Cheng AC, Rao SK, Tang E, Lam DS. Pachymetry assessment with Orbscan II in postoperative patients with myopic LASIK. J Refract Surg. 2006; 22:363–6.
crossref
16. Prisant O, Calderon N, Chastang P, et al. Reliability of pachymetric measurements using Orbscan after excimer refractive surgery. Ophthalmology. 2003; 110:511–5.
crossref
17. Yaylali V, Kaufman SC, Thompson HW. Corneal thickness measurements with the Orbscan Topography System and ultrasonic pachymetry. J Cataract Refract Surg. 1997; 23:1345–50.
crossref
18. Giessler S, Duncker GI. Orbscan pachymetry after LASIK is not reliable. J Refract Surg. 2001; 17:385–7.
crossref
19. Boscia F, La Tegola MG, Alessio G, Sborgia C. Accuracy of Orbscan optical pachymetry in corneas with haze. J Cataract Refract Surg. 2002; 28:253–8.
crossref
20. Fakhry MA, Artola A, Belda JI, et al. Comparison of corneal pachymetry using ultrasound and Orbscan II. J Cataract Refract Surg. 2002; 28:248–52.
crossref
21. von Jagow B, Kohnen T. Corneal architecture of femtosecond laser and microkeratome flaps imaged by anterior segment optical coherence tomography. J Cataract Refract Surg. 2009; 35:35–41.
crossref

Figure 1.
Mean and 95% CI of CCT measurement by Orbscan, Pentacam, Galilei and US in Group I (CI=confidence interval; CCT=central.thickness; US=Ultrasonic pachymetry).
jkos-51-320f1.tif
Figure 2.
Mean and 95% CI of CCT Measurement by Orbscan, Pentacam, Galilei and US in Group II (CI= confidence interval; CCT=central. thickness; US= Ultrasonic pachymetry).
jkos-51-320f2.tif
Figure 3.
Mean and 95% CI of CCT Measurement by Orbscan, Pentacam, Galilei and US in Group III. (CI= confidence interval; CCT=central. thickness; US=Ultrasonic pachymetry).
jkos-51-320f3.tif
Figure 4.
Bland-Altman plots of Galilei, Pentacam and Orbscan CCT readings against US measurement in Group I. The middle line is the mean and the lines on the side represent the upper and lower 95% limits of agreement (LoA).
jkos-51-320f4.tif
Figure 5.
Bland-Altman plots of Galilei, Pentacam and Orbscan CCT readings against US measurement in Group II. The middle line is the mean and the lines on the side represent the upper and lower 95% limits of agreement (LoA).
jkos-51-320f5.tif
Figure 6.
Bland-Altman plots of Galilei, Pentacam and Orbscan CCT readings against US measurement in Group III. The middle line is the mean and the lines on the side represent the upper and lower 95% limits of agreement (LoA).
jkos-51-320f6.tif
Table 1.
Characteristics of the study groups
  Group I Group II Group III§
Patients (eyes) 30 30 30
Sex (M:F) 5:10 2:13 2:13
Age (mean) 27.45±7.49 29.33±7.49 25.92±4.25
Manifest refractive error (SE) (D)* –4.50±1.72 –4.25±1.65 –4.19±1.61
SE Correction (D)*   –4.64±1.70 –4.51±1.64
Follow-up duration (Month)   5.80±4.33 11.23±7.26

* SE=spherical equivalent

I=unoperated eyes

II=eyes 1 to 6 months after Femtosecond laser-assisted LASIK

§ III=eyes 6 months or more after Femtosecond laser-assisted LASIK.

Table 2.
Corneal thickness measurements in normal and post-femtosecond laser-assisted LASIK eyes using orbscan II, pentacam, galilei and ultrasound pachymetry (Mean± SE, μm)
Group Orbscan Pentacam Galilei US* P value
I 535.8±40.7 544.5±30.7 554.2±30.4 541.8±31.3 0.202
II 440.3±40.3 478.8±28.6 488.0±30.1 474.7±32.1 .000
III§ 432.2±51.7 472.5±33.5 482.8±32.9 470.0±33.3 .000

* US=ultrasound

I=unoperated eyes

II=eyes 1 to 6 months after Femtosecond laser-assisted LASIK

§ III=eyes 6 months or more after Femtosecond laser-assisted LASIK

Corrected Orbscan value=raw Orbscan value ×0.95 of custom acoustic correction factor

p<0.05; one-way ANOVA was used.

Table 3.
Mean difference, limit of agreements and pearson correlation in corneal thickness measurements for galilei, pentacam, orbscan and ultrasound pachymetry in group I, II and III
Group and measurement Mean difference (μm) 95% Limit of agreements (μm) P Value Pearson correlation (R)(p<.001)
Group I        
   Galilei-US* 12.3 –10.2, 34.9 0.487 0.982
   Pentacam-US* 2.7 –19.9, 25.3 0.989 0.984
   Orbscan-US* –6.1 –28.6, 16.5 0.897 0.932
Group II        
   Galilei-US* 13.2 –9.0, 35.5 0.412 0.983
   Pentacam-US* 4.1 –18.2, 26.4 0.964 0.980
   Orbscan-US* –34.4 –56.7, −12.1 0.001 0.903
Group III§        
   Galilei-US* 12.8 –13.2, 38.9 0.573 0.972
   Pentacam-US* 2.4 –23.6, 28.5 0.995 0.977
   Orbscan-US* –37.8 –63.8, −11.7 0.001 0.918

* US=ultrasound

I=unoperated eyes

II=eyes 1 to 6 months after Femtosecond laser-assisted LASIK

§ III=eyes 6 months or more after Femtosecond laser-assisted LASIK

Corrected Orbscan value=raw Orbscan value ×0.95 of custom acoustic correction factor

p<0.05; one-way ANOVA and Scheffé multiple comparison test was used.