Abstract
Purpose
To report the change of anterior chamber parameters according to cataract severity after cataract surgery and to determine its relationship to the severity of cataract by using anterior segment optical coherence tomography.
Methods
We measured the anterior chamber parameters in 19 eyes of 14 patients before, 1 week after, and 1 month after cataract surgery by slit lamp-adapted optical coherence tomography (SL-OCT). The measured parameters were as follows : the anterior chamber depth (ACD), the angle-opening distance 250 μ m from the scleral spur (AOD250), the angle-opening distance 500 μ m from the scleral spur (AOD500), and the trabecular-iris angle (TIA). We analyzed the relationship between the severity of cataract and the change of the anterior chamber parameters.
Results
The ACD, AOD250, AOD500, and TIA increased significantly at postoperative 1 week ( P=0.000, 0.002, 0.005, 0.022) and 1 month ( P=0.000, 0.004, 0.001, 0.002). The preoperative parameters were negatively correlated with the differences between the postoperative 1 week and preoperative parameters (γ = -0.834, -0.591, -0.421, -0.826) and between postoperative 1 month and preoperative parameters (γ = -0.659, -0.700, -0.770, -0.821). The change of parameters at postoperative 1 week (by N P=0.959, 0.916, 0.824, 1.000, by C P=0.454, 0.665, 0.578, 0.578) and 1 month (by N P=0.858, 0.973, 0.959, 0.959, by C P=0.999, 0.207, 0.950, 0.981) were not significantly different according to the severity of cataract (N, C).
References
1. Steuhl KP, Marahrens P, Frohn C, Frohn A. Intraocular pressure and anterior chamber depth before and after extracapsular cataract extraction with posterior chamber lens implantation. Ophthalmic Surg. 1992; 23:233–7.
2. Koo BS, Chung J, Baek NH. The effect of extracapsular cataract extraction in patients with chronic angle-closure glaucoma combined with cataract. J Korean Ophthalmol Soc. 1996; 37:1045–53.
3. Kurimoto Y, Park M, Sakaue H, Kondo T. Changes in the anterior chamber configuration after small-incision cataract surgery with posterior chamber intraocular lens implantation. Am J Ophthalmol. 1997; 124:775–80.
4. Lee HJ, Chung SK, Baek NH. Changes of preoperative and postoperative anterior chamber angle in phacoemulsification and planned extracapsular cataract extraction. J Korean Ophthalmol Soc. 1998; 39:1170–5.
5. Pereira FA, Cronemberger S. Ultrasound biomicroscopic study of anterior segment changes after phacoemulsification and foldable intraocular lens implantation. Ophthalmology. 2003; 110:1799–806.
6. Nonaka A, Kondo T, Kikuchi M. . Angle widening and alteration of ciliary process configuration after cataract surgery for primary angle closure. Ophthalmology. 2006; 113:437–41.
7. Memarzadeh F, Tang M, Li Y. . Optical coherence tomography assessment of angle anatomy changes after cataract surgery. Am J Ophthalmol. 2007; 144:464–5.
8. Shibata T, Hockwin O, Weigelin E. . Biometry of the lens with respect to age and cataract morphology. Evaluation of Scheimpflug photos of the anterior segment. Klin Monatsbl Augenheilkd. 1984; 185:35–42.
9. Richard DW, Russell SR, Anderson DR. A method for improved biometry of the anterior chamber with a Scheimpflug technique. Invest Ophthalmol Vis Sci. 1988; 29:1826–35.
10. Shibata T, Sazuki K, Skamoto Y, Takahashi N. Quantitative chamber angle measurement utilizing image-processing techniques. Ophthalmic Res. 1990; 22:–S81. 4
11. Pavlin CJ, Harasiewiez K, Foster FS. Ultrasound biomicroscopy of anterior segment structures in normal and glaucomatous eyes. Am J Ophthalmol. 1992; 113:381–9.
12. Wirbelauer C, Gochmann R, Pham DT. Imaging of the anterior eye chamber with optical coherence tomography. Klin Monatsbl Augenheilkd. 2005; 222:856–62.
13. Radhakrishnan S, Goldsmith J, Huang D. . Comparison of optical coherence tomography and ultrasound biomicroscopy for detection of narrow anterior chamber angles. Arch Ophthalmol. 2005; 123:1053–9.
14. Müler M, Dahmen G, Pörksen E. . Anterior chamber angle measurement with optical coherence tomography: Intraobserver and interobserver variability. J Cataract Refract Surg. 2006; 32:1803–8.
15. Murphy GE. Long-term gonioscopy follow-up of eyes with posterior lens implants and no iridectomy. Ophthalmic Surg. 1986; 17:227–8.
17. Radhakrishnan S, Huang D, Smith SD. Optical coherence tomography imaging of anterior chamber angle. Ophthalmol Clin North Am. 2005; 18:375–81.
18. Lee JH, Park WC, Rho SH. The effects of pilocarpine on the anterior chamber depth and angle. J Korean Ophthalmic Soc. 1994; 35:572–9.
19. Arai M, Ohzuno I, Zako M. Anterior chamber depth after posterior chamber intraocular lens implantation. Acta Ophthalmol. 1994; 72:694–7.
20. Yoshida S, Hashiba H, Tsukuda M, Ohara Y. Significance of angle of intraocular lens haptics on anterior chamber depth. Jpn J Clin Ophthalmol. 1989; 43:173–6.
Table 1.
Variable | Preoperative | 1 week | 1 month | 1 week difference | 1 month difference |
---|---|---|---|---|---|
ACD (µm)∗ | 2670.32±653.66 | 3591.21±456.91 | 3735.63±605.34 | 920.89±698.45 | 1065.32±779.79 |
AOD250 (µm)† | 342.58±127.27 | 470.11±123.78 | 470.16±113.10 | 127.53±149.33 | 127.58±154.30 |
AOD500 (µm)‡ | 532.53±226.22 | 725.05±235.07 | 768.32±150.31 | 192.53±244.09 | 235.79±244.74 |
TIA (degrees)$ | 34.92±13.36 | 43.26±8.66 | 48.37±7.92 | 8.34±13.19 | 13.45±13.01 |
Table 2.
Variable | 1 week’s difference | 1 month’s difference | 1 week’s ratio | 1 month’s ratio |
---|---|---|---|---|
ACD (µm)† | -0.834 | -0.659 | -0.875 | -0.789 |
AOD250 (µm)‡ | -0.591 | -0.700 | -0.810 | -0.792 |
AOD500 (µm)§ | -0.421 | -0.770 | -0.638 | -0.817 |
TIA (degress)∏ | -0.826 | -0.821 | -0.855 | -0.894 |
Table 3.
Variable | Group 1 (N2, N3) | Group 2 (N4, N5) | P value∗ |
---|---|---|---|
ACD (µm)† | 873.60±711.48 | 973.44±722.66 | 0.959 |
AOD250 (µm)‡ | 131.76±126.13 | 122.89±179.53 | 0.916 |
AOD500 (µm)§ | 157.00±273.45 | 232.00±215.87 | 0.824 |
TIA (degress)∏ | 9.00±12.97 | 7.61±14.18 | 1.000 |
Table 4.
Variable | Group 1 (N2, N3) | Group 2 (N4, N5) | P value∗ |
---|---|---|---|
ACD (µm)†† | 1129.30±965.33 | 994.22±555.88 | 0.858 |
AOD250 (µm)‡ | 121.60±120.22 | 134.22±192.92 | 0.973 |
AOD500 (µm)§ | 230.10±195.42 | 242.11±302.86 | 0.959 |
TIA (degress)∏ | 14.80±11.70 | 11.94±14.89 | 0.959 |
Table 5.
Variable | Group 1 (C2, C3) | Group 2 (C4, C5) | P value∗ |
---|---|---|---|
ACD (µm)† | 916.15±625.20 | 931.17±904.39 | 0.454 |
AOD250 (µm)‡ | 123.54±176.17 | 136.17±75.21 | 0.665 |
AOD500 (µm)§ | 179.46±286.75 | 220.83±125.44 | 0.578 |
TIA (degress)∏ | 7.58±15.66 | 9.98±5.76 | 0.578 |