Abstract
Purpose
To determine the rate of congruence and to standardize assessment of US (ultrasound) phantom images with the use of an ATS-539 multipurpose phantom for US equipment currently utilized in Korea
Materials and Methods
US phantom images were scanned with a 3.0-5.0 MHz convex transducer and were digitized by use of an analogue-digital converter. Members of a committee with consent evaluated the US phantom images from 108 types of ultrasound equipment. The dead zone, vertical and horizontal measurement, axial/lateral resolution, focal zone, sensitivity, functional resolution and gray scale/dynamic range were evaluated. Congruence or incongruence of ultrasound equipment was determined based on the results of dead zone, axial/lateral resolution and gray scale/dynamic range measurements. Other factors were evaluated for the possibility as criteria with the use of the Mann-Whitney U test and receiver operator characteristic (ROC) curve analysis.
Results
The dead zone, axial/lateral resolution and gray scale/dynamic range were 91.7%, 94.4% and 76.9%, respectively, for suitable US equipment. Considering all three factors, 78 types of ultrasound equipment were passed. The congruence rate of focal zone and functional resolution were 62.4% and 69.3% of the US equipment, respectively.
Conclusion
Of the US equipment, 72.2% of the equipment was acceptable based on the dead zone, axial/lateral resolution, and gray scale/dynamic range measurements as determined with the use of an ATS-539 phantom. Focal zone and 8 mm-functional resolution can be useful as a standard in the assessment of a US phantom image.