Abstract
Purpose
To assess the influence of stent design [interlacing (type A) vs. crossing method (type B)] on neointimal hyperplasia using a self-expandable nitinol stent, which crosses the side branches of the abdominal aorta.
Materials and Methods
In seven mongrel dogs, nitinol stents with type A and B intersections were placed in the abdominal aorta across the main branches: 4 Niti-Ms and 3 Niti-Ds. Two months after the stent placement, a DSA was performed for a stent patency evaluation, followed by the extraction of the aortas. The degree of neointimal formation along the wire was evaluated by calculating the area ratio of intimal hyperplasia (type A vs. B). A Student's t-test was employed to investigate the differences in the neointimal hyperplasia between blood types A and B.
Results
The total number of wire intersections overlain at the ostia branch ostia was 23 for type A and 36 for type B. The area ratio of the neointimal hyperplasia, for a given area, was 29.09±10.82% (type A) and 13.80±6.94% (type B) (p < 0.0001)]. Furthermore, the area ratios of the neointimal hyperplasia per area of stent-wire in the given area were 138.38±10.84% (type A), 87.58±7.36% (type B) (p = 0.0002).