Abstract
Purpose
To examine whether the ACR phantom could be used in quality standards for magnetic resonance imaging (MRI) instruments in Korea.
Materials and Methods
We conducted the phantom test using the ACR MRI phantom in 20 MRI instruments currently used in Korea. According to ACR criteria, we acquired the phantom images which were then assessed by the following seven tests: geometric accuracy, high spatial resolution, slice thickness accuracy, slice position accuracy, image intensity uniformity, percent signal ghosting, and low contrast object detectability. The phantom images were interpreted by three experienced radiologists according to ACR criteria. Then, we examined the failure rate of each test and evaluated the inter-observer variation in the measurements and test failure.
Results
The failure rate of each test could be broken into the following components: geometric accuracy (11-21%), high contrast spatial resolution (10-15%), slice thickness accuracy (6-22%), slice position accuracy (5-17%), image intensity uniformity (6%), percent signal ghosting (16%), and low contrast object detectability (8-10%). In this series, all the failure rates were less than 30%. In addition, no inter-observer variation was seen in the measurements and test failure.