Journal List > Tuberc Respir Dis > v.62(4) > 1001075

Choi, Choi, Park, Kang, and Yoo: Clinical Value of a Desktop Spirometer (HI-801) for Spirometry Screening

Abstract

Background

A national health care initiative recommends routine spirometry screening of all smokers over age 45 or patients with respiratory symptoms. In response to the recommendation, new, simple, and inexpensive desktop spirometers for the purpose of promoting widespread spirometric screening were marketed. The performance of these spirometers was evaluated in vivo testing with healthy subjects. However, the clinical setting allows spirometric assessment of various pathologic combinations of flow and volume.

Objective

The aim of this study was to compare the accuracy of a desktop spirometer to a standard laboratory spirometer, in a clinical setting with pathologic pulmonary function.

Method

In a health check-up center, where screening pulmonary funct test was performed using the HI-801 spirometer. Subjects who revealed the ventilation defect in screening spirometry, performed the spirometry again using the standard Vmax spectra 22d spirometer in a tertiary care hospital pulmonary function laboratory. Pulmonary function test with both spirometer was performed according to the guidelines of the American Thoracic Society.

Results

109 patients were enrolled. Pulmonary function measurements (FVC, FEV1, PEFR, FEF25%-75%) from the HI-801 correlated closely (r=0.94, 0.93, 0.81, 0.84, respectively) with those performed with the Vmax spectra 22d and showed the good limits of agreement and differences between the 2 devices; FVC +0.35 L, FEV1 +0.16 L, PEFR +1.85 L/s, FEF25%-75% -0.13 L/s. With the exception of FEV1, FEF25%-75%, these differences were significant(p<0.05) but small.

Conclusion

The HI-801 spirometer is comparable to the standard laboratory spirometer, Vmax spectra 22d, with high accurary for FEV1 and FVC and acceptable differences for clinical use.

Figures and Tables

Figure 1
Bivariate correlation analysis between lung function measurements made with the HI-801 and Vmax spirometer.
(A) FVC, (B) FEV1, (C) FEF25%-75%, (D) PEFR.
FEV1: forced expiratory volume in 1 second; FVC: forced vital capacity; PEFR: peak expiratory flow rate; FEF25%-75%: forced expiratory flow at 50% of the forced vital capacity. (p<0.01 for all comparisons).
trd-62-276-g001
Figure 2
Differences versus mean values for lung function measurements made with HI-801 and Vmax 22d spirometers.
The solid lines represent the mean differences. The dotted lines represent the upper and lower limits of agreement (±1.96SD) (In Bland Altman plot).
FEV1: forced expiratory volume in 1 second; FVC: forced vital capacity; PEFR: peak expiratory flow rate; FEF25%-75%: forced expiratory flow at 50% of the forced vital capacity.
trd-62-276-g002
Table 1
Comparison of 2 spirometers
trd-62-276-i001
Table 2
Demographics of the participants
trd-62-276-i002

Mean±Standard Deviation

Table 3
Differences of ventilation defect between HI-801 and Vmax 22d spirometry
trd-62-276-i003
Table 4
Comparison of measurements from HI-801 and Vmax 22d
trd-62-276-i004

Mean±Standard Deviation

*Calculated as mean Vmax 22d reading minus mean HI-801 reading, p<0.01 using bivariate correlation analysis (r=pearson correlation coefficient), p<0.05 using paired t-test

FEV1: forced expiratory volume in 1 second,; FVC: forced vital capacity; PEFR: peak expiratory flow rate; FEF25%-75%: forced expiratory flow at 50% of the forced vital capacity.

References

1. Ferguson GT, Enright PL, Buist AS, Higgins MW. Office spirometry for lung health assessment in adults: a consensus statement from the National Lung Health Education Program. Chest. 2000. 117:1146–1161.
2. The World Health Report. 2002: reducing risks, promoting healthy life. 2002. Geneva, Switzerland: World Health Organization;Available from: http://www.who.int/whr/2002.
3. Wise RA, Kanner RE, Lindgren P, Connett JE, Altose MD, Enright PL, et al. The effect of smoking intervention and an inhaled bronchodilator on airways reactivity in COPD: the Lung Health Study. Chest. 2003. 124:449–458.
4. Anthonisen NR, Connett JE, Kiley JP, Altose MD, Bailey WC, Buist AS, et al. Effects of smoking intervention and the use of an inhaled anticholinergic bronchodilator on the rate of decline of FEV1. JAMA. 1994. 272:1497–1505.
5. Miller MR, Hankinson J, Brusasco V, Burgos F, Casaburi R, Coates A, et al. Standardisation of spirometry. Eur Respir J. 2005. 26:319–338.
6. American Thoracic Society. Standardization of spirometry, 1994 update. Am J Respir Crit Care Med. 1995. 152:1107–1136.
7. Korhonen H, Remes ST, Kannisto S. Hand-held turbine spirometer: agreement with the conventional spirometer at baseline and after exercise. Pediatr Allergy Immunol. 2005. 16:254–257.
8. Maree DM, Videler EA, Hallauer M, Pieper CH, Bolliger CT. Comparison of a new desktop spirometer (Diagnosa) with a laboratory spirometer. Respiration. 2001. 68:400–404.
9. Swart F, Schuurmans MM, Heydenreich JC, Pieper CH, Bolliger CT. Comparison of a new desktop spirometer (Spirospec) with a laboratory spirometer in a respiratory out-patient clinic. Respir Care. 2003. 48:591–595.
10. Rebuck DA, Hanania NA, D'Urzo AD, Chapman KR. The accuracy of a handheld portable spirometer. Chest. 1996. 109:152–157.
11. HI-801® Operator's Manual. Version 2.7. 2004. Tokyo, Japan: CHEST M.I., Inc.
12. Vmax® Operator's Manual. 2003. Yorba Linda, CA: Sensor Medics Corporation.
13. Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986. 1:307–310.
14. Schoh RJ, Fero LJ, Shapiro H, Aslor JP, Kaelin OJ, Rollins DR, Petty TL. Performance of a new screening spirometer at a community health fair. Respir Care. 2002. 47:1150–1157.
15. Robert ED, Katherine LV, Jennifer C, Shawn DA. Spirometry in primary care setting: Influence on clinical diagnosis and management of airflow obstruction. Chest. 2005. 128:2443–2447.
16. Izbicki G, Abboud S, Jordan P, Perruchoud AP, Bolliger CT. A comparison of a new transtelephonic portable spirometer with a laboratory spirometer. Eur Respir J. 1999. 14:209–213.