Journal List > Korean J Pain > v.24(4) > 1159353

Cho, Kim, Kang, Park, Kim, Lim, Kim, and Woo: A Study to Compare the Radiation Absorbed Dose of the C-arm Fluoroscopic Modes

Abstract

Background

Although many clinicians know about the reducing effects of the pulsed and low-dose modes for fluoroscopic radiation when performing interventional procedures, few studies have quantified the reduction of radiation-absorbed doses (RADs). The aim of this study is to compare how much the RADs from a fluoroscopy are reduced according to the C-arm fluoroscopic modes used.

Methods

We measured the RADs in the C-arm fluoroscopic modes including 'conventional mode', 'pulsed mode', 'low-dose mode', and 'pulsed + low-dose mode'. Clinical imaging conditions were simulated using a lead apron instead of a patient. According to each mode, one experimenter radiographed the lead apron, which was on the table, consecutively 5 times on the AP views. We regarded this as one set and a total of 10 sets were done according to each mode. Cumulative exposure time, RADs, peak X-ray energy, and current, which were viewed on the monitor, were recorded.

Results

Pulsed, low-dose, and pulsed + low-dose modes showed significantly decreased RADs by 32%, 57%, and 83% compared to the conventional mode. The mean cumulative exposure time was significantly lower in the pulsed and pulsed + low-dose modes than in the conventional mode. All modes had pretty much the same peak X-ray energy. The mean current was significantly lower in the low-dose and pulsed + low-dose modes than in the conventional mode.

Conclusions

The use of the pulsed and low-dose modes together significantly reduced the RADs compared to the conventional mode. Therefore, the proper use of the fluoroscopy and its C-arm modes will reduce the radiation exposure of patients and clinicians.

INTRODUCTION

Fluoroscopy is a machine that generates X-ray images and presents them continuously as visible images during a diagnostic or interventional procedure [1]. C-arm fluoroscopy is frequently utilized and absolutely necessary in modern pain medicine [2]. Although fluoroscopy enhances the patients' safety and increases the accuracy, it also exposes clinicians to the cumulative doses of ionizing radiation [3]. The adverse effects of ionizing radiation on human bodies include skin diseases, thyroid cancer, brain tumors, cataracts and so on [4]. Such effects are largely divided into two types. The "early effects" are acute radiation lethality, local tissue damage on the skin or gonads, hematologic effects, and cytogenetic effects, and the "late effects" are radiation-induced malignancies such as leukemia and other forms of cancer, deleterious local tissue effects, chromosomal toxicity, and/or cataract formation [5]. Therefore, for the safety of the patients and clinicians it is essential and mandatory to reduce the radiation exposure as far as possible. Available patient dose reduction technologies include low fluoroscopy dose rate settings, low frame rate pulsed fluoroscopy, spectral beam filtration, and use of increased X-ray beam energy [6]. Although pulsed and low-dose fluoroscopy has the potential to greatly reduce the radiation exposure of clinicians involved in interventional procedures, few studies have quantified the reduction in radiation-absorbed doses (RADs) according to the C-arm modes. The purpose of this study is to compare how much the RADs are reduced when using fluoroscopy in the C-arm modes.

MATERIALS AND METHODS

We measured the cumulative exposure time, RADs, peak X-ray energy (kVp), current (mA) in the C-arm fluoroscopic modes including the 'conventional mode', 'pulsed mode', 'low-dose mode', and 'pulsed + low-dose mode'. A pulsed fluoroscopic mode of 15 frames per second was used. The C-arm fluoroscopy was the OEC® fluoroscopic unit (9800 Plus, Salt LakeCity, Utah) and the automatic brightness control (ABC) that manipulates the peak X-ray energy and the current of the fluoroscopic X-ray beam that provides the appropriate image contrast and brightness on the viewing monitor were used in the study. Clinical imaging conditions that are typical in interventional procedures were simulated using a 0.85 mm lead equivalent thick apron instead of a patient. According to each mode, one experimenter radiographed the lead apron, which was on the table, 5 times consecutively on the anteroposterior view and tried to press the X-ray on switch for the same time intervals (Fig. 1). We regarded this as one set and a total 10 sets were done according to each mode. The cumulative exposure time, RADs, peak X-ray energy, and current that were viewed on the monitor were recorded. Data were analyzed by one-way ANOVA followed by Turkey's test for multiple comparisons. Analytical results were considered statistically significant if the P value was less than 0.05.

RESULTS

The mean cumulative exposure time, RADs, peak X-ray energy, and current according to the C-arm modes are presented in Table 1.

1. Time

The mean cumulative exposure time was significantly lower in the pulsed (1.6 ± 0.5) and pulsed + low-dose modes (1.0 ± 0.2) than in the conventional mode (2.7 ± 0.1). The low-dose mode and conventional mode had the same mean cumulative exposure time (2.7 ± 0.1) (Table 1, Fig. 2).

2. RADs and RADs/Time

The mean RADs was significantly lower in the pulsed (5.6 ± 1.8), low-dose (3.5 ± 0.2), and pulsed + low-dose modes (1.4 ± 0.3) than in the conventional mode (8.2 ± 0.4). The mean RADs/Time was significantly lower in the low-dose (1.2) and pulsed + low-dose modes (1.4) than in the conventional mode (3.1). The mean RADs/Time in pulsed mode (3.5) was similar to that of the conventional mode (3.1) (Table 1, Fig. 2).

3. Peak X-ray energy and Current

All modes were pretty much the same for the peak X-ray energy. However, the mean peak X-ray energy was higher in the pulsed + low-dose mode (83.9 ± 0.3) than in any other modes and the mean peak X-ray energy was lower in the conventional mode (80.0) than in any of the other modes. The mean current was significantly lower in the low-dose (1.1) and pulsed + low-dose modes (1.1) than in the conventional mode (2.8 ± 0.4). The mean current in the pulsed mode (3.1 ± 0.1) was similar to that of the conventional mode (2.8 ± 0.4) (Table 1, Fig. 2, and 3).

DISCUSSION

Modern fluoroscopes typically offer two modes of fluoroscopy: either the X-ray beam is continuously on, or the X-ray beam is pulsed on for only a few strobe-like pulses of X-rays each second [7]. Manufacturers commonly offer clinicians a choice of several different fluoroscopic pulse rates ranging from a minimum of 2 or 3 frames/s, to a maximum of 30 frames/s (usually called continuous fluoroscopy by most manufacturers) [8].
Pulsed fluoroscopy can always decrease the radiation dose. In this study, the 15 frames per second pulsed mode showed a significantly decreased exposure time by 41% and decreased RADs by 32% compared to the conventional mode. The reason for this is that manufacturers increase the duration or the amplitude or power of each fluoroscopic radiation pulse as the fluoroscopic pulse rate is decreased. They do this to retain at least reasonable image quality [9]. With the low-dose mode, the exposure is reduced through the adjustment of the peak X-ray energy and current to create the X-ray beam [3]. In this study, the low-dose mode showed a significantly decreased current by 60% and decreased RADs by 57% and slightly increased peak X-ray energy by 4% compared to the conventional mode. The pulsed + low-dose mode showed significantly decreased RADs by 83% and decreased exposure time by 63% compared to the conventional mode. Conclusively, using the pulsed mode and low-dose modes together showed significantly decreased RADs than in any other mode.
Pulsed fluoroscopy has 2 disadvantages. On a real-time fluoroscopic image, as the pulse rate is lowered, the cumulative radiation dose decreases, and the real-time fluoroscopic image becomes more noisy or grainy [9]. According to the C-arm modes, four fluoroscopic images blocking the lumbar medial branch nerves were shown in Fig. 4. The fluoroscopic image was darker and grainier in the pulsed mode (B) than in the conventional mode (A) because the increased current applied to the X-ray tube produces more X-rays, and the more X-rays that strike the image intensifier, the darker the image is [10]. The fluoroscopic image was brighter with less contrast between the different tissues in the low-dose mode (C) than in the conventional mode (A), thereby reducing the image detail. In our experience, generally, a low-dose fluoroscopic image is a slightly better detailed image than the 15 frames per second pulsed fluoroscopic image. However, the X-ray dose and image quality are interrelated and the image may differ due to the patients weight, fluoroscopic pulse rate, peak X-ray energy, current and X-ray tube-patient-image intensifier distance [11,12]. A trade-off between image quality and radiation exposure is inevitable, however, supportive techniques are considered to provide optimal image quality while reducing radiation exposure to a level as low as reasonably achievable.
Collimation is helpful in optimizing the image quality because the ABC mode attempts to optimize the image quality by taking into account the exposure needed across the entire field of exposure. Useful application of collimation can exclude areas of greatly varying radiodensity to improve the image quality by reducing the range of densities included in the field [6]. The clarity of small structures, or image detail, can be improved by lowering the peak X-ray energy and by reducing the distance between the patient and the image intensifier [10].
Second, with very slow pulse rates, motion such as swallowing or peristalsis becomes jerky [9]. However, continuous fluoroscopy is not the only alternative. The human visual system requires a certain of amount of time (0.2 s) to "integrate" an image. This integration time suggests that there may be little image-quality reward from continuous fluoroscopy, because the human visual system will produce strobelike images for interpretation, regardless if the X-ray beam is on continuously or not [13]. Pulsed fluoroscopy can also provide better temporal resolution for moving objects than continuous fluoroscopy since pulsed fluoroscopy provides strobelike "stop-motion" images at a variable number of pulses per second [14].
This study has limitations such as not measuring the biological effects of the RADs (BERADs) and only measuring the RADs itself. However, since RAD is proportional to BERAD, we can predict the BERADs on the body according to the C-arm fluoroscopic modes.
In conclusion, the use of the pulsed and low-dose modes together significantly reduced the RADs compared to the conventional mode. Therefore, the proper use of the fluoroscopy and its C-arm modes will reduce the radiation exposure of patients and clinicians. Because of the reduced image quality with the pulsed and low-dose settings, however, an ideal way to reduce the RADs would be to use these modes intermittently during a procedure. When the clinicians do not need detailed image quality such as a medial branch block, they can use the pulsed and low-dose modes together. If clinicians need images with more details, using the low-dose mode is another alternative. Supportive techniques such as collimation, lowering the peak X-ray energy, reducing the distance between the patient and the image intensifier are considered to provide the optimal image quality. Continuous or conventional modes can be used during image setup for a procedure using distance and shielding to protect against radiation exposure.

References

1. Raj PP, Lou L, Erdine S, Staats PS, Waldman SD, Racz G, et al. Interventional pain management: Image-guided procedures. 2008. 2nd ed. Philadelphia: Saunders Elsevier;p. 11.
2. Kim TW, Jung JH, Jeon HJ, Yoon KB, Yoon DM. Radiation exposure to physicians during interventional pain procedures. Korean J Pain. 2010; 23:24–27. PMID: 20552069.
crossref
3. Goodman BS, Carnel CT, Mallempati S, Agarwal P. Reduction in average fluoroscopic exposure times for interventional spinal procedures through the use of pulsed and low-dose image settings. Am J Phys Med Rehabil. 2011; [Epub ahead of print].
crossref
4. Cousins C, Sharp C. Medical interventional procedures-reducing the radiation risks. Clin Radiol. 2004; 59:468–473. PMID: 15145716.
crossref
5. Mroz TE, Yamashita T, Davros WJ, Lieberman IH. Radiation exposure to the surgeon and the patient during kyphoplasty. J Spinal Disord Tech. 2008; 21:96–100. PMID: 18391712.
crossref
6. Miller DL, Vañó E, Bartal G, Balter S, Dixon R, Padovani R, et al. Occupational radiation protection in interventional radiology: a joint guideline of the Cardiovascular and Interventional Radiology Society of Europe and the Society of Interventional Radiology. Cardiovasc Intervent Radiol. 2010; 33:230–239. PMID: 20020300.
crossref
7. Brown PH, Thomas RD, Silberberg PJ, Johnson LM. Optimization of a fluoroscope to reduce radiation exposure in pediatric imaging. Pediatr Radiol. 2000; 30:229–235. PMID: 10789900.
crossref
8. Cohen MD. Can we use pulsed fluoroscopy to decrease the radiation dose during video fluoroscopic feeding studies in children? Clin Radiol. 2009; 64:70–73. PMID: 19070700.
crossref
9. Cohen MD. Optimizing the use of pulsed fluoroscopy to reduce radiation exposure to children. J Am Coll Radiol. 2008; 5:205–209. PMID: 18312969.
crossref
10. Rathmell JP. Atlas of image-guided intervention in regional anesthesia and pain medicine. 2006. Philadelphia: Lippincott Williams & Wilkins;p. 9–15.
11. Uppot RN, Sahani DV, Hahn PF, Kalra MK, Saini SS, Mueller PR. Effect of obesity on image quality: fifteen-year longitudinal study for evaluation of dictated radiology reports. Radiology. 2006; 240:435–439. PMID: 16801372.
crossref
12. Wagner LK, Archer BR, Cohen AM. Management of patient skin dose in fluoroscopically guided interventional procedures. J Vasc Interv Radiol. 2000; 11:25–33. PMID: 10693710.
crossref
13. Aufrichtig R, Xue P, Thomas CW, Gilmore GC, Wilson DL. Perceptual comparison of pulsed and continuous fluoroscopy. Med Phys. 1994; 21:245–256. PMID: 8177157.
crossref
14. Brown PH, Silberberg PJ, Thomas RD, Strife JL, Towbin RB. A multihospital survey of radiation exposure and image quality in pediatric fluoroscopy. Pediatr Radiol. 2000; 30:236–242. PMID: 10789901.
crossref
Fig. 1
The lead apron (0.85 mm lead equivalent thickness) was on the table and the X-ray tube was positioned below the apron and the detector above the apron.
kjpain-24-199-g001
Fig. 2
Comparison of time, radiation-absorbed doses (RADs), mean RADs/mean time and current (mA) among modes by graphs. Pulsed fluoroscopic mode of 15 frames per second was used. *P < 0.05. Unit is expressed as second in Time, mRADs/cm2 in RADs, mRADs/cm2 · second in RADs/Time, and mA in Current.
kjpain-24-199-g002
Fig. 3
Comparison of peak X-ray energy (kVp) among modes by graphs. Pulsed fluoroscopic mode of 15 frames per second was used. *P < 0.05.
kjpain-24-199-g003
Fig. 4
According to C-arm modes, four fluoroscopic images blocking lumbar medial branch nerves. (A) Convention mode. (B) Pulsed mode. (C) Low-dose mode. (D) Pulsed + low-dose mode.
kjpain-24-199-g004
Table 1
Mean Cumulative Exposure Time, Radiation-absorbed Doses (RADs), Peak X-ray Energy (kVp), Current (mA) according to C-arm Modes
kjpain-24-199-i001

Data are presented as means ± SDs. Analytical results were considered statistically significant if the *P value was less than 0.05. Pulsed fluoroscopic mode of 15 frames per second was used.

TOOLS
Similar articles