Traditionally, anesthesiologists have recommended performing auscultation of breath and heart sounds with a precordial or esophageal stethoscope during anesthesia. The stethoscope is generally considered the most important single device used for monitoring. Unfortunately, many anesthesiologists neglect the perioperative use of a stethoscope, even though it provides a wealth of information about the circulatory system, heart, and lungs. Although point-of-care ultrasound appears to offer greater diagnostic value in some situations, it is premature to dismiss the stethoscope, and its diminished perioperative role should be reconsidered. Recent technological advances have allowed changes in the roles of these modalities to help attending anesthesiologists perform easy assessments using clinically meaningful information from auscultation. The phonocardiopulmography (PCPG), a real-time visual acoustic monitor, is a rediscovered and renovated technique that allows non-invasive continuous heart-lung function monitoring. However, the novel role of the PCPG as a continuous acoustic monitor, particularly for S1 and S2 heart sounds and their respiratory variations, remains relatively unexplored. This device may convey information about cardiac contractility, cardiac output, and fluid responsiveness. Furthermore, continuous lung sound visualization has potential as a monitoring modality during anesthesia, including during bronchospasm and endobronchial intubation. With its ease of use and intuitive mechanism, the PCPG can be applied during general anesthesia and follow-up in both post-anesthesia and intensive care unit settings, particularly for the detection of acute adverse events during the perioperative period.
Auscultation is an essential bedside diagnostic method used in modern medicine. The practice of auscultation was first described by Hippocrates in roughly 460 to 370 BC [1]. This rudimentary form of auscultation was conducted by applying a physician’s ear to the patient’s chest to listen to sounds from within the body. Although Harvey formally described heart sounds in 1628, immediate auscultation was employed less frequently than palpation or percussion. In 1816, Rene Laennec invented the first stethoscope, comprising a perforated wooden cylinder, a funnel-shaped plug, and a stopper. The stethoscope gained popularity and became an essential medical tool in the 19th and 20th centuries. Today, however, auscultation is increasingly being replaced by evolving technologies such as echocardiography, which provides qualitative and quantitative information with greater ease. Consequently, the role of auscultation is being redefined, transitioning from a comprehensive diagnostic tool, to a practical, inexpensive, and rapid screening method that can aid general clinicians in making timely decisions at the bedside [2].
The PCG, which provides a graphic record of heart sounds, was developed by Einthoven and Geluc in 1894 to provide objective information on auscultation [3]. Through the 1930s and the 1940s, monitoring and recording equipment for PCG was developed, and standardization began in 1950. A PCG system manufactured in 1965 was used in at least one of the Gemini project crewed spaceflights to monitor the heartbeat of astronauts during flight (Fig. 1). The PCG records heart sounds using a microphone, and the resulting acoustic signal is transformed into a visual waveform. With recent advances in digital techniques, PCG can now characterize heart sounds and murmurs based on their acoustic patterns, components, timing, and frequency. Currently, however, the PCG is not commonly utilized by cardiologists, and physicians are unaware of its existence. Although this technology has advanced and remained available for several decades, it has largely been supplanted by echocardiography. Nevertheless, several recent studies have evaluated the efficacy of PCG in the diagnosis of structural heart disease [4], ventricular dysfunction [5], and heart failure [6,7]. Recently, commercialized PCG with electrocardiography (ECG) recording devices have been introduced for outpatient screening [8]. Furthermore, the development of machine learning and artificial intelligence has dramatically improved their utility [9].
According to the American Society of Anesthesiologists standards for basic intraoperative monitoring, every patient receiving general anesthesia should undergo continuously evaluation using qualitative clinical signs, including auscultation of breath sounds, to ensure adequate ventilation. In earlier literature, it was recommended that anesthesiologists auscultate breath and heart sounds with a precordial or esophageal stethoscope during routine, pediatric, thoracic, regional, and monitored anesthesia care [10,11]. The anesthesia stethoscope has also been identified as the most important single device used for monitoring [12]. Although the precise impact of stethoscope use during anesthesia on patient outcomes remains unclear, a theoretical analysis of critical anesthesia incidents utilizing esophageal stethoscopes suggested that vigilant continuous auscultation could prevent 54% of such events [13]. Despite these recommendations, perioperative auscultation is often neglected, and has been largely replaced by capnography and pulse oximetry. The use of esophageal or precordial stethoscopes during anesthesia markedly decreased from 70–87% in the mid-1980s to 28% in the late-1990s [14]. In some reports, the stethoscope is considered inferior to point-of-care ultrasound for differentiating tracheal and bronchial intubation [15,16]. Although recent data are lacking, a minority of anesthesia staff and trainees may actively auscultate heart and breath sounds with a stethoscope during anesthesia. Unfortunately, many anesthesiologists neglect the use of a stethoscope perioperatively, even though it provides a wealth of information about the circulatory system, heart, and lungs [17]. Still, many anesthesiologists do not support abandoning the stethoscope and acknowledge its utility in a variety of clinical situations, when properly used. However, it is important to recognize the low sensitivity and specificity of auscultation, even in the hands of experienced clinicians [15]. To overcome this limitation, digitalized stethoscopes have been introduced and applied clinically in procedural sedation [18], preoperative patient evaluation [19], and postoperative care using wearable patch forms [20]. Although intraoperative PCG is listed in the Korean healthcare claim book, there are currently no commercially available PCG devices, except those used in experimental equipment. Recently, attention has been paid to changing the role of this technique in helping attending anesthesiologists perform assessments easily using clinically meaningful information from auscultation.
Normal heart sounds consist of the S1 (atrioventricular valve closure, marking the onset of the ventricular systolic phase) and S2 (aortic and pulmonary valve closure, marking the end of the ventricular systolic phase) components. Physiological and pathological changes in the heart sounds can predict cardiovascular abnormalities. Cardiac murmurs can be easily detected by auscultation, and the PCG offers a more intuitive and objective assessment of the operating room (Fig. 1). Perioperative PCG, a simple and noninvasive procedure, warrants further investigation as a potential method for monitoring cardiac performance.
The usefulness of intraoperative PCG as a non-invasive monitor for cardiac performance has previously been investigated. Pharmacological treatments can influence heart sounds; for example, inotropic agents (β-adrenergic drugs) have strong impacts on S1, while systemic vascular resistance–modifying drugs (afterload) have a relatively prominent effect on S2. In canine models, both anesthetic depression and dopamine stimulation of the cardiovascular system were clearly demonstrated using PCG [21]. Changes in the amplitude of the S1 heart sound were found to closely correlate with changes in left ventricular pressure. In a swine model, the amplitude of the S1 heart sound was significantly correlated with cardiac contractility during the administration of dobutamine and esmolol [22]. In contrast, changes in vascular resistance caused by the administration of phenylephrine and nicardipine did not show any significant correlation. In beagles, the amplitude of the S1 heart sound correlated significantly with changes in cardiac output under various conditions [23]. The heart rate measured using an esophageal stethoscope closely matched the ECG values, showing a strong positive correlation. In humans, S1 amplitude correlated positively with the cardiac index in patients undergoing laparoscopic surgery [24]. In 17 patients undergoing liver transplantation, the S1 heart sound was found to be predominantly modulated by ephedrine/esmolol, whereas the S2 heart sound was mainly modified by phenylephrine/nicardipine, but not vice versa [25].
Proportional changes in heart sounds were observed when the blood pressure was changed using inotropes or volatile anesthetics [26]. The relationship between S1/S2 and the change in blood pressure showed a positive correlation in the inotropic experimental model. In the volatile anesthetic experiment, the heart sounds decreased as the minimum alveolar concentration increased. Using an esophageal stethoscope and wireless Bluetooth transmission, the ratio of S1 to S2 showed a greater correlation with systolic blood pressure than S1 alone [27]. It has previously been reported that the beat-to-beat measured S1–S2 interval using an esophageal stethoscope was closely correlated with the corresponding pulse pressure [28]. Continuous monitoring of the intraoperative S1–S2 interval has the potential to act as an indirect estimator of beat-to-beat arterial pulse pressure, and respiratory variations of the S1–S2 interval appears to mirror pulse pressure variation with moderate accuracy (Fig. 2), In 30 liver transplantation recipients, the beat-to-beat monitoring of the S1–S2 interval and its respiratory variation was shown to predict fluid responsiveness, and may be useful as a non-invasive hemodynamic index for fluid management [29]. These results indicate that intraoperative PCG may be useful in detecting cardiovascular changes, such as alterations in cardiac output, contractility, and afterload, which commonly occur in patients receiving anesthesia.
Normal lung sounds consist of inhaled and exhaled components. Changes in the pitch and frequency of the lung sounds can be applied to detect abnormalities in the respiratory system. In patients receiving mechanical ventilation, the lung sound signals typically show different morphological patterns depending on the ventilator mode (Fig. 3). Preoperative meticulous auscultation of the lungs helps to determine whether wheezing or rhonchi should be initiated intraoperatively.
Several reports have previously described the features of lung sounds under certain clinical conditions. In the porcine model, the severity and distribution of oleic acid-induced lung injury can be assessed by acoustic changes recorded over the trachea and chest wall [30]. The changes in dependent lung sound transmission had high within-subject correlations with venous admixture and dynamic compliance. Kato et al. [31] developed a system that allows the visualization of breath sounds that display the left and right lung sound simultaneously. During advancement of the tracheal tube, changes in the shape of the visualized acoustic signals using the visual stethoscope appeared before the changes in breath sounds were detected by auscultation. The bilateral breath sounds disappeared when the tip of the tracheal tube advanced beyond the carina. In pediatric patients, the visual stethoscope detects changes in breath sounds and unilateral breath sounds as the tracheal tube advances [32].
The spectral analysis of breath sounds obtained from the esophageal stethoscope can be a useful non-invasive respiratory monitor for intratracheal secretion accumulation [33]. Detecting thick intratracheal secretions with power ratio of 80–200 Hz and 300–400 Hz showed the highest discriminating ability. Song et al. [34] reported that intraoperative bronchospasms can be detected early through real-time acoustic visualization using a digitalized esophageal stethoscope. The visualization of lung sounds facilitates the rapid identification of expiratory wheezing and the abnormal spectrogram patterns characteristic of bronchospasms. To meet these clinical needs, a six-channel acoustic monitoring system for use in surgical patients during perioperative period was developed, and various types of microphone designs were evaluated [35]. The displayed information included time and frequency domain signals, and tools to aid in the detection of adverse events. We hypothesized that such an arrangement may help anesthesiologists to recognize clinical conditions such as wheezing, bronchospasm, endobronchial intubation, and apnea. Respiratory acoustic monitoring (Acoustic Respiration Rate [RRa®], Masimo Inc.) is a commercialized device used to monitor the respiratory rate in patients recovering from anesthesia. Under general anesthesia with a laryngeal mask airway and spontaneous ventilation, RRa® rapidly detects changes in the respiratory rate with minimal bias [36].
The use of PCG as a monitoring modality in the field of anesthesia has several limitations. First, the signal quality is influenced by ambient background noise. Although the operating room is quiet and still, acoustic signals can be inherently contaminated by endogenous and exogenous noise. Adequate noise processing filters are therefore required to ensure robust signal quality [37]. Secondly, optimal positioning of the microphone is important for obtaining a proper auscultation signal. Well-established auscultation positions and landmarks are recommended for precordial probes. In contrast, placement depth greatly affects the quality of auscultation sounds using an esophageal probe. A depth of 28–32 cm is recommended for clinical use; S1, S2, and inspiratory and expiratory sounds have a high amplitude in that range [38]. Third, the sample sizes of recent studies have not been large enough to establish robust cutoff values for each parameter. Further studies with larger sample sizes are therefore required to generalize these results.
Recently, the application of machine learning and deep learning to PCG data has increased. For example, constructive probabilistic neural network and radial basis function network have been reported to classify a broad range of tracheal-bronchial breath sounds [39]. Integration of the convolutional Fourier modules further ensures the robust performance of PCG segmentation, even amidst heart murmurs and background noise, thereby significantly contributing to the advancement of cardiac diagnostics [40]. Shin et al. [41] previously proposed temporal convolutional network-based feature extraction and information fusion to integrate asynchronously acquired PCG recordings at the patient level. The proposed framework accommodates variable-length, multi-channel PCG recordings, making it an accurate and flexible solution for detecting pediatric cardiac abnormalities, particularly in low-resource settings.
The novel role of the PCG as a continuous acoustic monitor, particularly for S1, S2 and their respiratory variations remain relatively unexplored. Furthermore, continuous lung sound visualization also has potential monitoring modality in the field of anesthesia. Technically, digitized auscultation data share the same signal source and can be separated into heart and lung sound components using specific frequency filters. In this respect, we propose the concept of phonocardiopulmography (PCPG), a technique which enables the simultaneous visualization of both the heart and lung sounds (Fig. 4). Further larger scale follow-up data are required to define the role of the PCPG, a visual acoustic monitor, as a cost-effective bedside tool to identify adverse events to supplement physical examination, which is commonly neglected in busy clinical settings.
Innovation should always remain an important aim as we strive to improve the safety and reliability of anesthesiology. Although point-of-care ultrasound appears to offer greater diagnostic value in various situations, it is premature to dismiss the stethoscope, and its diminished perioperative role should be reconsidered. Visual acoustic monitoring (PCPG) is a novel technique that allows non-invasive continuous heart-lung function monitoring. Owing to its ease of use and relatively low cost, the PCPG can be applied during general anesthesia and follow-up in both post-anesthesia and intensive care unit settings, particularly for the detection of acute adverse events during the perioperative period.
Notes
ACKNOWLEDGMENTS
The author would like to thank Woo Jin Kim (Ph.D. student) for his help in preparing the manuscript.
REFERENCES
1. Hanna IR, Silverman ME. A history of cardiac auscultation and some of its contributors. Am J Cardiol. 2002; 90:259–67. DOI: 10.1016/s0002-9149(02)02465-7. PMID: 12173582.

2. Tavel ME. Cardiac auscultation: a glorious past--and it does have a future! Circulation. 2006; 113:1255–9. DOI: 10.1161/circulationaha.105.591149. PMID: 16520426.
3. Montinari MR, Minelli S. The first 200 years of cardiac auscultation and future perspectives. J Multidiscip Healthc. 2019; 12:183–9. DOI: 10.2147/JMDH.S193904. PMID: 30881010.
4. Papunen I, Ylänen K, Lundqvist O, Porkholm M, Rahkonen O, Mecklin M, et al. Automated analysis of heart sound signals in screening for structural heart disease in children. Eur J Pediatr. 2024; 183:4951–8. DOI: 10.1007/s00431-024-05773-3. PMID: 39304593.

5. Shapiro M, Moyers B, Marcus GM, Gerber IL, McKeown BH, Vessey JC, et al. Diagnostic characteristics of combining phonocardiographic third heart sound and systolic time intervals for the prediction of left ventricular dysfunction. J Card Fail. 2007; 13:18–24. DOI: 10.1016/j.cardfail.2006.09.009. PMID: 17338999.

6. Cao L, Guo X, Liao K, Qin J, Zheng Y. A comprehensive nomogram integrating phonocardiogram and echocardiogram features for the diagnosis of heart failure with preserved ejection fraction. Clin Cardiol. 2024; 47:e70022. DOI: 10.1002/clc.70022. PMID: 39465895.

7. Butler J, Brown M, Prokocimer P, Humphries AC, Pope S, Wright O, et al. The role of cardiac acoustic biomarkers in monitoring patients with heart failure: a systematic literature review. ESC Heart Fail. 2025; 12:980–97. DOI: 10.1002/ehf2.15075. PMID: 39294891.

8. Wen YN, Lee AP, Fang F, Jin CN, Yu CM. Beyond auscultation: acoustic cardiography in clinical practice. Int J Cardiol. 2014; 172:548–60. DOI: 10.1016/j.ijcard.2013.12.298. PMID: 24529949.

9. Fuadah YN, Lim KM. Advances in cardiovascular signal analysis with future directions: a review of machine learning and deep learning models for cardiovascular disease classification based on ECG, PCG, and PPG signals. Biomed Eng Lett. 2025; 15:619–60. DOI: 10.1007/s13534-025-00473-9. PMID: 40625553.

10. Calvery RK, Scheller M. Anesthesia as a specialty: past, present and future. In: Clinical anesthesia. 2nd ed. Edited by Barash PG, Cullen BF, Stoelting RK: Philadelphia, J. B. Lippincott. 1992, p 947.
11. Zelcer J, White PF. Monitored anesthesia care. In: Anesthesia. 3rd ed. Edited by Miller RD: New York, Churchill Livingstone. 1990, p 1327.
12. Perrino AC Jr, Feldman JM, Barash PG. Noninvasive cardiovascular monitoring. In: Monitoring in anesthesia. 3rd ed. Edited by Saidman LJ, Smith NT: Boston, Butterworth-Heinemann. 1993, pp 136-8.
13. Klepper ID, Webb RK, Van der Walt JH, Ludbrook GL, Cockings J. The Australian Incident Monitoring Study. The stethoscope: applications and limitations--an analysis of 2000 incident reports. Anaesth Intensive Care. 1993; 21:575–8. DOI: 10.1177/0310057x9302100514. PMID: 8273878.

14. Prielipp RC, Kelly JS, Roy RC. Use of esophageal or precordial stethoscopes by anesthesia providers: are we listening to our patients? J Clin Anesth. 1995; 7:367–72. DOI: 10.1016/0952-8180(95)00027-f. PMID: 7576670.

15. Ramsingh D, Frank E, Haughton R, Schilling J, Gimenez KM, Banh E, et al. Auscultation versus point-of-care ultrasound to determine endotracheal versus bronchial intubation: a diagnostic accuracy study. Anesthesiology. 2016; 124:1012–20. DOI: 10.1097/ALN.0000000000001073. PMID: 26950708.
16. Isono S, Sandberg WS, Jiang Y. Do you believe what you see or what you hear? Ultrasound versus stethoscope for perioperative clinicians. Anesthesiology. 2016; 124:989–91. DOI: 10.1097/ALN.0000000000001074. PMID: 26950709.
17. Jablons ML. The value of the stethoscope in the era of ultrasound. Anesthesiology. 2016; 125:1248–9. DOI: 10.1097/aln.0000000000001373.

18. Kishimoto N, Yamamoto T, Sakagami R, Seo K. Respiratory monitoring during intravenous sedation using a digital auscultation device. J Dent Anesth Pain Med. 2025; 25:215–7. DOI: 10.17245/jdapm.2025.25.3.215. PMID: 40521427.

19. Ahuja V, Kini P, Nileshwar A. Evaluation of the electronic stethoscope (FONODOC) as a cardiac screening tool during the preoperative evaluation of children. Indian J Anaesth. 2022; 66:625–30. DOI: 10.4103/ija.ija_305_22. PMID: 36388445.

20. Mohamed N, Kim HS, Mohamed M, Kang KM, Kim SH, Kim JG. Tablet-based wearable patch sensor design for continuous cardiovascular system monitoring in postoperative settings. Biosensors (Basel). 2023; 13:615. DOI: 10.3390/bios13060615. PMID: 37366980.

21. Hansen PB, Luisada AA, Miletich DJ, Albrecht RF. Phonocardiography as a monitor of cardiac performance during anesthesia. Anesth Analg. 1989; 68:385–7. DOI: 10.1213/00000539-198903000-00037. PMID: 2919778.

22. Park YS, Kim HS, Lee SA, Hwang GS, Jung W, Moon B, et al. Correlations between heart sound components and hemodynamic variables. Sci Rep. 2024; 14:8602. DOI: 10.1038/s41598-024-59362-3. PMID: 38615106.

23. Duck Shin Y, Hoon Yim K, Hi Park S, Wook Jeon Y, Ho Bae J, Soo Lee T, et al. The correlation between the first heart sound and cardiac output as measured by using digital esophageal stethoscope under anaesthesia. Pak J Med Sci. 2014; 30:276–81. PMID: 24772126.
24. Bickel A, Eitan A, Melnik D, Weiss A, Gavrieli N, Kniaz D, et al. The use of pneumoperitoneum during laparoscopic surgery as a model to study pathophysiologic phenomena: the correlation of cardiac functionality with computerized acoustic indices--preliminary data. J Laparoendosc Adv Surg Tech A. 2012; 22:349–54. DOI: 10.1089/lap.2011.0360. PMID: 22416808.

25. Moon YJ, Kim SH, Park YS, Kim JM, Hwang GS. Quantitative analysis of an intraoperative digitalized esophageal heart sound signal to speculate on perturbed cardiovascular function. J Clin Med. 2019; 8:715. DOI: 10.3390/jcm8050715. PMID: 31137509.

26. Park SH, Shin YD, Bae JH, Kwon EJ, Lee TS, Shin JY, et al. Heart sounds analysis via esophageal stethoscope system in beagles. J Clin Monit Comput. 2013; 27:535–9. DOI: 10.1007/s10877-013-9459-0. PMID: 23529343.

27. Hoon Lim K, Duck Shin Y, Hi Park S, Ho Bae J, Jae Lee H, Jung Kim S, et al. Correlation of blood pressure and the ratio of S1 to S2 as measured by esophageal stethoscope and wireless bluetooth transmission. Pak J Med Sci. 2013; 29:1023–7. DOI: 10.12669/pjms.294.3639. PMID: 24353680.
28. Park YS, Moon YJ, Kim SH, Kim JM, Song JG, Hwang GS. Beat-to-beat tracking of pulse pressure and its respiratory variation using heart sound signal in patients undergoing liver transplantation. J Clin Med. 2019; 8:593. DOI: 10.3390/jcm8050593. PMID: 31052236.

29. Kim SH, Moon YJ, Kim JW, Song JG, Hwang GS. Prediction of fluid responsiveness by a non-invasive respiratory systolic time interval variation using heart sound signals in recipients undergoing liver transplantation. Transplant Proc. 2017; 49:1082–1086. DOI: 10.1016/j.transproceed.2017.03.032. PMID: 28583532.

30. Räsänen J, Gavriely N. Detection of porcine oleic acid-induced acute lung injury using pulmonary acoustics. J Appl Physiol (1985). 2002; 93:51–7. DOI: 10.1152/japplphysiol.01238.2001. PMID: 12070185.
31. Kato H, Suzuki A, Nakajima Y, Makino H, Sanjo Y, Nakai T, et al. A visual stethoscope to detect the position of the tracheal tube. Anesth Analg. 2009; 109:1836–42. DOI: 10.1213/ane.0b013e3181bb4967. PMID: 19923511.

32. Kimura T, Suzuki A, Mimuro S, Makino H, Sato S. Bronchial intubation could be detected by the visual stethoscope techniques in pediatric patients. Paediatr Anaesth. 2012; 22:1185–90. DOI: 10.1111/pan.12010. PMID: 22928677.

33. Moon YJ, Bechtel AJ, Kim SH, Kim JW, Thiele RH, Blank RS. Detection of intratracheal accumulation of thick secretions by using continuous monitoring of respiratory acoustic spectrum: a preliminary analysis. J Clin Monit Comput. 2020; 34:763–70. DOI: 10.1007/s10877-019-00359-z. PMID: 31327100.

34. Song J, Moon B, Kim HS, Seo WY, Kim WJ, Kim SH. Diagnostic value of visual stethoscopes for detecting bronchospasm in a patient under general anesthesia - a case report. Anesth Pain Med (Seoul). 2025; 20:277–82. DOI: 10.17085/apm.24194. PMID: 40792372.
35. Jafarian K, Amineslami M, Hassani K, Navidbakhsh M, Lahiji MN, Doyle DJ. A multi-channel acoustics monitor for perioperative respiratory monitoring: preliminary data. J Clin Monit Comput. 2016; 30:107–18. DOI: 10.1007/s10877-015-9693-8. PMID: 25869899.

36. Atkins JH, Mandel JE. Performance of Masimo rainbow acoustic monitoring for tracking changing respiratory rates under laryngeal mask airway general anesthesia for surgical procedures in the operating room: a prospective observational study. Anesth Analg. 2014; 119:1307–14. DOI: 10.1213/ANE.0000000000000362. PMID: 25025587.
37. Mohamed N, Kim HS, Kang KM, Mohamed M, Kim SH, Kim JG. Heart and lung sound measurement using an esophageal stethoscope with adaptive noise cancellation. Sensors (Basel). 2021; 21:6757. DOI: 10.3390/s21206757. PMID: 34695968.

38. Manecke GR Jr, Poppers PJ. Esophageal stethoscope placement depth: its effect on heart and lung sound monitoring during general anesthesia. Anesth Analg. 1998; 86:1276–9. DOI: 10.1097/00000539-199806000-00027. PMID: 9620519.
39. Folland R, Hines E, Dutta R, Boilot P, Morgan D. Comparison of neural network predictors in the classification of tracheal-bronchial breath sounds by respiratory auscultation. Artif Intell Med. 2004; 31:211–20. DOI: 10.1016/j.artmed.2004.01.004. PMID: 15302087.

40. Park C, Shin K, Seo J, Lim H, Kim GH, Seo WY, et al. Enhancement of phonocardiogram segmentation using convolutional neural networks with Fourier transform module. Biomed Eng Lett. 2025; 15:401–13. DOI: 10.1007/s13534-025-00458-8. PMID: 40026887.

41. Shin JM, Park S, Shin K, Seo WY, Kim HS, Kim DK, et al. Temporal convolutional neural network-based feature extraction and asynchronous channel information fusion method for heart abnormality detection in phonocardiograms. Comput Methods Programs Biomed. 2025; 269:108871. DOI: 10.1016/j.cmpb.2025.108871. PMID: 40460727.

Fig. 1.
Representative plot of the phonocardiogram of a patient with ventricular septal defect before (left) and after correction surgery (right). Note that the typical murmur of the ventricular septal defect disappeared after surgery.
Fig. 2.
The upper panel displays the changes in stroke volume variation (gray) and acoustic variability index (green) during the operation. The blue shadow at the start of surgery and red shadow during active bleeding indicate an euvolemic and hypovolemic status, respectively. The lower panel displays the actual shape of arterial blood pressure and phonocardiogram during the start of surgery (blue) and active bleeding (red). The Y-axis of the lower panel represents blood pressure (mmHg) and heart sound (arbitrary units).
Fig. 3.
Representative plot of the lung sound signal according to mechanical ventilator setting, including volume control ventilation (left) and pressure control ventilation (right). The upper panel represents the raw respiratory signal (blue) segmented by each breath cycle (red). The lower panel represents time-averaged (red dot) and raw signal (thin line) during each respiratory cycle. Green and red arrows represent the width of inhale and exhale of respiration, respectively. The plateau height indicates the maximum signal amplitude during the breathing-in phase. The plateau width indicates the duration of the stable breathing-in period, measured in seconds. The full width at half maximum (FWHM) indicates the width of the inspiration curve at half of its maximum height. P_ratio indicates the peak position within FWHM. (t_Ex-t_Ins)/t_Res represents the normalized time interval between inspiration and expiration peaks.
Fig. 4.
(A) Block diagram of the phonocardiopulmography (PCPG) system. Heart and lung sounds are collected by the signal acquisition module, and transferred to signal processing module. Processed signals and calculated indices are displayed in real-time. (B) Components of the PCPG system and the conceptual image of a visual acoustic monitor (Signal House, Co, Ltd), which displays both the heart and lung sounds and their derivative parameters.