Journal List > J Korean Fract Soc > v.26(4) > 1037953

Kwak, Kim, Young, Joh, and Hwang: Breakage of Reamer during Tibia Intramedullary Nailing - A Case Report -

Abstract

The reamer crack, followed by breakage at its distal part occurred during intramedullary nailing of tibial shaft fracture. The broken reamer was trapped in the intramedullary canal, making it very difficult to pull out. We successfully extracted the broken reamer by retrograde impaction through the fracture site and completed intramedullary nailing procedure. Thus, we present this case with a review of the literature.

Figures and Tables

Fig. 1
Preoperative anteroposterior and lateral radiographs of 37-year-old male patient. The patient was injured from slipping down with external rotation force.
jkfs-26-333-g001
Fig. 2
Fluoroscopic radiograph of distal end of the broken reamer.
jkfs-26-333-g002
Fig. 3
We successfully pulled out the broken reamer using a guidewire in a retrograde manner through the additional incision over the fracture site.
jkfs-26-333-g003
Fig. 4
Broken reamer shaft after removal: Distal and proximal end of broken reamer were twisted and elongated.
jkfs-26-333-g004
Fig. 5
Open reduction with bone clamp and internal fixation with intramedullary nail and interlocking screws were performed.
jkfs-26-333-g005
Fig. 6
Postoperative anteroposterior and lateral radiographs of the tibia.
jkfs-26-333-g006

References

1. Bhandari M, Guyatt GH, Swiontkowski MF, et al. Surgeons' preferences for the operative treatment of fractures of the tibial shaft. An international survey. J Bone Joint Surg Am. 2001; 83-A:1746–1752.
2. Bisbinas I, Ho YM, Learmonth DJ. Reamer breakage in the femoral medulla during total knee arthroplasty: should reamers have a finite life? J Arthroplasty. 2004; 19:501–503.
3. Coles CP, Gross M. Closed tibial shaft fractures: management and treatment complications. A review of the prospective literature. Can J Surg. 2000; 43:256–262.
4. Darowish M, Gorczyca JT. Catastrophic intramedullary fragmentation of a tinel reamer. J Trauma. 2009; 67:E36–E40.
crossref
5. Hooper GJ, Keddell RG, Penny ID. Conservative management or closed nailing for tibial shaft fractures. A randomised prospective trial. J Bone Joint Surg Br. 1991; 73:83–85.
crossref
6. Hwang SK, Yoo JM. Complication related to interlocking nailing of the tibia fracture. J Korean Soc Fract. 1995; 8:823–829.
crossref
7. Moreland JR, Marder R, Anspach WE Jr. The window technique for the removal of broken femoral stems in total hip replacement. Clin Orthop Relat Res. 1986; (212):245–249.
crossref
8. Noh JH, Yang BK, Park JT, Je MS. Breakage of core reamer during anterior cruciate ligament reconstruction -A case report-. J Korean Arthrosc Soc. 2010; 14:33–35.
9. Pape HC, Giannoudis P. The biological and physiological effects of intramedullary reaming. J Bone Joint Surg Br. 2007; 89:1421–1426.
crossref
TOOLS