Journal List > J Korean Orthop Assoc > v.45(2) > 1012975

Kim and Kim: Radiation Exposure to the Orthopaedic Surgeon during Fracture Surgery

Abstract

Purpose

To assess the risk of radiation exposure by measuring the equivalent dose during fracture surgery.

Materials and Methods

Two orthopedic trauma surgeons were enrolled in this study from April to June 2009. The equivalent dose was measured using thermoluminiscent dosimeters placed at the left (in and out of the lead apron).

Results

The equivalent dose measured from the apron over the 3 month period were 5.22 mSv, 4.34 mSv(outside), and 1.83 mSv and 0.71 mSv(inside). The rate of radiation reduction was 64.9% and 83.6% respectively.

Conclusion

The estimated annual equivalent dose outside the apron was close to or higher than the maximum limit of radiation exposure. These findings highlight the need for surgeons to be more cautious about radiation exposure during fracture management and take active steps to minimize radiation exposure, such as wearing a radio-protective apron.

Figures and Tables

Figure 1
Photograph showing the location of TLD (thermo luminescent dosimeter). One TLD is placed at the left chest (in and out of the apron). The radio-protective apron is a half apron.
jkoa-45-107-g001
Figure 2
C-arm fluoroscopy with x-ray tube and image intensifier. (A) Typical C-arm fluoroscopy. (B) Inverted C-arm fluoroscopy.
jkoa-45-107-g002
Figure 3
Stray radiation fields for lateral projection. (A) 100 cm and (B) 150 cm above the floor (Reprinted, with permission, from reference 25).
jkoa-45-107-g003
Table 1
Annual Recommended Limits for Occupational and Nonoccupational Radiation Exposure
jkoa-45-107-i001

*The international Commission on Radiological Protection recommends a maximum of 20 mSv/yr; the National Council on Radiation Protection and Measurements recommends a maximum of 50 mSv/yr.

Table 2
The Equivalent Dose (mSv) Exposure Measured with TLD
jkoa-45-107-i002

TLD, thermo-luminescent dosimeter.

Table 3
Estimates of Exposure Time during Surgical Procedure
jkoa-45-107-i003

MIPO, minimal invasive plate osteosynthesis; ORIF, open reduction and internal fixation; m, minute; s, second.

Table 4
Distribution of Annual Radiation Exposure of Korean Doctors in 2007
jkoa-45-107-i004

References

1. Oh CW, Kim JJ, Byun YS, et al. Minimally invasive plate osteosynthesis of subtrochanteric femur fractures with a locking plate: a prospective series of 20 fractures. Arch Orthop Trauma Surg. 2009. 129:1659–1665.
crossref
2. Apivatthakakul T, Chiewcharntanakit S. Minimally invasive plate osteosynthesis (MIPO) in the treatment of the femoral shaft fracture where intramedullary nailing is not indicated. Int Orthop. 2008. 33:1119–1126.
crossref
3. Zhiquan A, Bingfang Z, Yeming W, Chi Z, Peiyan H. Minimally invasive plating osteosynthesis (MIPO) of middle and distal third humeral shaft fractures. J Orthop Trauma. 2007. 21:628–633.
crossref
4. Hasenboehler E, Rikli D, Babst R. Locking compression plate with minimally invasive plate osteosynthesis in diaphyseal and distal tibial fracture: a retrospective study of 32 patients. Injury. 2007. 38:365–370.
crossref
5. Hazarika S, Chakravarthy J, Cooper J. Minimally invasive locking plate osteosynthesis for fractures of the distal tibia--results in 20 patients. Injury. 2006. 37:877–887.
crossref
6. Perren SM. The technology of minimally invasive percutaneous osteosynthesis (MIPO). Injury. 2002. 33:Suppl 1. VI–VII.
7. Krettek C, Muller M, Miclau T. Evolution of minimally invasive plate osteosynthesis (MIPO) in the femur. Injury. 2001. 32:Suppl 3. SC14–SC23.
crossref
8. Maxon HR, Thomas SR, Saenger EL, Buncher CR, Kereiakes JG. Ionizing irradiation and the induction of clinically significant disease in the human thyroid gland. Am J Med. 1977. 63:967–978.
crossref
9. Merriam GR Jr., Focht EF. A clinical study of radiation cataracts and the relationship to dose. Am J Roentgenol Radium Ther Nucl Med. 1957. 77:759–785.
10. 1990 Recommendations of the International Commission on Radiological Protection. Ann ICRP. 1991. 21:1–201.
11. Ron E. Cancer risks from medical radiation. Health Phys. 2003. 85:47–59.
crossref
12. Bross ID, Ball M, Falen S. A dosage response curve for the one rad range: adult risks from diagnostic radiation. Am J Public Health. 1979. 69:130–136.
crossref
13. Korea Food and Drug Administration. 2007 Report occupational radiation exposure in diagnostic radiology in korea. 2008. Seoul: Food and Drug Administration.
14. Kempf I, Grosse A, Beck G. Closed locked intramedullary nailing. Its application to comminuted fractures of the femur. J Bone Joint Surg Am. 1985. 67:709–720.
crossref
15. Levin PE, Schoen RW Jr, Browner BD. Radiation exposure to the surgeon during closed interlocking intramedullary nailing. J Bone Joint Surg Am. 1987. 69:761–766.
crossref
16. Müller LP, Suffner J, Wenda K, Mohr W, Rommens PM. Radiation exposure to the hands and the thyroid of the surgeon during intramedullary nailing. Injury. 1998. 29:461–468.
crossref
17. Madan S, Blakeway C. Radiation exposure to surgeon and patient in intramedullary nailing of the lower limb. Injury. 2002. 33:723–727.
crossref
18. Devalia KL, Peter VK, Madanur MA, Braithwaite IJ. Exposure of the thyroid to radiation during routine orthopaedic procedures. Acta Orthop Belg. 2006. 72:615–620.
19. Barry TP. Radiation exposure to an orthopedic surgeon. Clin Orthop Relat Res. 1984. (182):160–164.
crossref
20. Goldstone KE, Wright IH, Cohen B. Radiation exposure to the hands of orthopaedic surgeons during procedures under fluoroscopic X-ray control. Br J Radiol. 1993. 66:899–901.
crossref
21. Fuchs M, Schmid A, Eiteljörge T, Modler M, Sturmer KM. Exposure of the surgeon to radiation during surgery. Int Orthop. 1998. 22:153–156.
crossref
22. Dosch JC, Dupuis M, Beck G. [Radiation measurements during interlocking nailing]. Hefte Unfallheilkd. 1983. 161:36–38.
23. Mehlman CT, DiPasquale TG. Radiation exposure to the orthopaedic surgical team during fluoroscopy: "how far away is far enough?". J Orthop Trauma. 1997. 11:392–398.
crossref
24. Tremains MR, Georgiadis GM, Dennis MJ. Radiation exposure with use of the inverted-c-arm technique in upper-extremity surgery. J Bone Joint Surg Am. 2001. 83-A:674–678.
crossref
25. Balter S. Stray radiation in the cardiac catheterisation laboratory. Radiat Prot Dosimetry. 2001. 94:183–188.
crossref
26. Jackson G, Brennan PC. Radio-protective aprons during radiological examinations of the thorax: An optimum strategy. Radiation Protection Dosimetry. 2006. 121:391–394.
crossref
27. Blattert TR, Fill UA, Kunz E, Panzer W, Weckbach A, Regulla DF. Skill dependence of radiation exposure for the orthopaedic surgeon during interlocking nailing of long-bone shaft fractures: a clinical study. Arch Orthop Trauma Surg. 2004. 124:659–664.
crossref
TOOLS
Similar articles