<?xml version="1.0" encoding="UTF-8"?>
<!DOCTYPE article PUBLIC "-//NLM//DTD JATS (Z39.96) Journal Publishing DTD v1.1 20151215//EN" "JATS-journalpublishing1.dtd">
<article xml:lang="EN" article-type="research-article">

<front>
<journal-meta>
<journal-id journal-id-type="nlm-ta">Healthc Inform Res</journal-id>
<journal-id journal-id-type="publisher-id">HIR</journal-id>
<journal-title-group>
<journal-title>Healthcare Informatics Research</journal-title>
</journal-title-group>
<issn pub-type="ppub">2093-3681</issn>
<issn pub-type="epub">2093-369X</issn>
<publisher>
<publisher-name>Korean Society of Medical Informatics</publisher-name>
</publisher>
</journal-meta>

<article-meta>
<article-id pub-id-type="doi">10.4258/hir.2019.25.4.274</article-id>
<article-categories>
<subj-group subj-group-type="heading">
<subject>Original Article</subject>
</subj-group>
</article-categories>
<title-group>
<article-title>Effect of Watch-Type Haptic Metronome on the Quality of Cardiopulmonary Resuscitation: A Simulation Study</article-title>
</title-group>

<contrib-group>

<contrib contrib-type="author">
<contrib-id contrib-id-type="orcid" authenticated="true">https://orcid.org/0000-0001-9244-604X</contrib-id>
<name>
<surname>Choi</surname>
<given-names>Boram</given-names>
</name>
<xref ref-type="aff" rid="A1">1</xref>
</contrib>

<contrib contrib-type="author">
<contrib-id contrib-id-type="orcid" authenticated="true">https://orcid.org/0000-0002-8149-1205</contrib-id>
<name>
<surname>Kim</surname>
<given-names>Taerim</given-names>
</name>
<xref ref-type="aff" rid="A1">1</xref>
</contrib>

<contrib contrib-type="author">
<contrib-id contrib-id-type="orcid" authenticated="true">https://orcid.org/0000-0002-4320-2388</contrib-id>
<name>
<surname>Yoon</surname>
<given-names>Sun Young</given-names>
</name>
<xref ref-type="aff" rid="A1">1</xref>
</contrib>

<contrib contrib-type="author">
<contrib-id contrib-id-type="orcid" authenticated="true">https://orcid.org/0000-0003-3696-3813</contrib-id>
<name>
<surname>Yoo</surname>
<given-names>Jun Sang</given-names>
</name>
<xref ref-type="aff" rid="A2">2</xref>
</contrib>

<contrib contrib-type="author">
<contrib-id contrib-id-type="orcid" authenticated="true">https://orcid.org/0000-0001-8196-0879</contrib-id>
<name>
<surname>Won</surname>
<given-names>Ho-Jeong</given-names>
</name>
<xref ref-type="aff" rid="A3">3</xref>
</contrib>

<contrib contrib-type="author">
<contrib-id contrib-id-type="orcid" authenticated="true">https://orcid.org/0000-0002-0865-2236</contrib-id>
<name>
<surname>Kim</surname>
<given-names>Kyunga</given-names>
</name>
<xref ref-type="aff" rid="A3">3</xref>
</contrib>

<contrib contrib-type="author">
<contrib-id contrib-id-type="orcid" authenticated="true">https://orcid.org/0000-0002-6324-1448</contrib-id>
<name>
<surname>Kang</surname>
<given-names>Eun Jin</given-names>
</name>
<xref ref-type="aff" rid="A1">1</xref>
</contrib>

<contrib contrib-type="author">
<contrib-id contrib-id-type="orcid" authenticated="true">https://orcid.org/0000-0002-1297-9813</contrib-id>
<name>
<surname>Yoon</surname>
<given-names>Hee</given-names>
</name>
<xref ref-type="aff" rid="A1">1</xref>
</contrib>

<contrib contrib-type="author">
<contrib-id contrib-id-type="orcid" authenticated="true">https://orcid.org/0000-0002-1352-3009</contrib-id>
<name>
<surname>Hwang</surname>
<given-names>Sung Yeon</given-names>
</name>
<xref ref-type="aff" rid="A1">1</xref>
</contrib>

<contrib contrib-type="author">
<contrib-id contrib-id-type="orcid" authenticated="true">https://orcid.org/0000-0001-9657-1040</contrib-id>
<name>
<surname>Shin</surname>
<given-names>Tae Gun</given-names>
</name>
<xref ref-type="aff" rid="A1">1</xref>
</contrib>

<contrib contrib-type="author">
<contrib-id contrib-id-type="orcid" authenticated="true">https://orcid.org/0000-0001-6645-3684</contrib-id>
<name>
<surname>Sim</surname>
<given-names>Min Seob</given-names>
</name>
<xref ref-type="aff" rid="A1">1</xref>
</contrib>

<contrib contrib-type="author" corresp="yes">
<contrib-id contrib-id-type="orcid" authenticated="true">https://orcid.org/0000-0002-2778-2992</contrib-id>
<name>
<surname>Cha</surname>
<given-names>Won Chul</given-names>
</name>
<xref ref-type="aff" rid="A1">1</xref>
<xref ref-type="aff" rid="A2">2</xref>
</contrib>

</contrib-group>

<aff id="A1"><label>1</label>Department of Emergency Medicine, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, <country>Korea</country>.</aff>
<aff id="A2"><label>2</label>Department of Digital Health, Samsung Advanced Institute for Health Sciences and Technology, Sungkyunkwan University, Seoul, <country>Korea</country>.</aff>
<aff id="A3"><label>3</label>Biostatistics and Clinical Epidemiology Center, Samsung Medical Center, Seoul, <country>Korea</country>.</aff>

<author-notes>
<corresp>Corresponding Author: Won Chul Cha. Department of Digital Health, Samsung Advanced Institute for Health Sciences and Technology, Sungkyunkwan University, 81 Irwon-ro, Gangnam-gu, Seoul 06351, Korea. Tel: +82-2-3410-2053, <email>wc.cha@samsung.com</email></corresp>
</author-notes>

<pub-date pub-type="ppub">
<month>10</month>
<year>2019</year>
</pub-date>
<pub-date pub-type="epub">
<day>31</day>
<month>10</month>
<year>2019</year>
</pub-date>
<volume>25</volume>
<issue>4</issue>
<fpage>274</fpage>
<lpage>282</lpage>

<history>
<date date-type="received">
<day>13</day>
<month>07</month>
<year>2019</year>
</date>
<date date-type="rev-recd">
<day>15</day>
<month>09</month>
<year>2019</year>
</date>
<date date-type="accepted">
<day>18</day>
<month>10</month>
<year>2019</year>
</date>
</history>

<permissions>
<copyright-statement>&#x00A9; 2019 The Korean Society of Medical Informatics</copyright-statement>
<copyright-year>2019</copyright-year>
<copyright-holder>The Korean Society of Medical Informatics</copyright-holder>
<license license-type="open-access" xlink:href="http://creativecommons.org/licenses/by-nc/4.0/">
<license-p>This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (<ext-link ext-link-type="uri" xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="http://creativecommons.org/licenses/by-nc/4.0/">http://creativecommons.org/licenses/by-nc/4.0/</ext-link>) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.</license-p>
</license>
</permissions>

<abstract>
<sec>
<title>Objectives</title>
<p>The aim of this study was to test the applicability of haptic feedback using a smartwatch to the delivery of cardiac compression (CC) by professional healthcare providers.</p>
</sec>
<sec>
<title>Methods</title>
<p>A prospective, randomized, controlled, case-crossover, standardized simulation study of 20 medical professionals was conducted. The participants were randomly assigned into haptic-first and non-haptic-first groups. The primary outcome was an adequate rate of 100&#x2013;120/min of CC. The secondary outcome was a comparison of CC rate and adequate duration between the good and bad performance groups.</p>
</sec>
<sec>
<title>Results</title>
<p>The mean interval between CCs and the number of haptic and non-haptic feedback-assisted CCs with an adequate duration were insignificant. In the subgroup analysis, both the good and bad performance groups showed a significant difference in the mean CC interval between the haptic and non-haptic feedback-assisted CC groups&#x2014;good: haptic feedback-assisted (0.57&#x2013;0.06) vs. non-haptic feedback-assisted (0.54&#x2013;0.03), <italic>p</italic> &#x003C; 0.001; bad: haptic feedback-assisted (0.57&#x2013;0.07) vs. non-haptic feedback-assisted (0.58&#x2013;0.18), <italic>p</italic> = 0.005&#x2014;and the adequate chest compression number showed significant differences&#x2014; good: haptic feedback-assisted (1,597/75.1%) vs. non-haptic feedback-assisted (1,951/92.2%), <italic>p</italic> &#x003C; 0.001; bad: haptic feedbackassisted (1,341/63.5%) vs. non-haptic feedback-assisted (523/25.4%), <italic>p</italic> &#x003C; 0.001.</p>
</sec>
<sec>
<title>Conclusions</title>
<p>A smartwatch cardiopulmonary resuscitation feedback system could not improve rescuers' CC rate. According to our subgroup analysis, participants might be aided by the device to increase the percentage of adequate compressions after one minute.</p>
</sec>
</abstract>

<kwd-group>
<kwd>Smartphone</kwd>
<kwd>Cardiopulmonary Resuscitation</kwd>
<kwd>Heart Massage</kwd>
<kwd>Feedback</kwd>
<kwd>Simulation Training</kwd>
</kwd-group>

</article-meta>
</front>

<body>

<sec sec-type="intro">
<title>I. Introduction</title>
<p>The survival rate from cardiac arrest remains poor despite ongoing efforts to improve the &#x2018;chain of survival&#x2019; over the past 50 years [<xref ref-type="bibr" rid="B1">1</xref><xref ref-type="bibr" rid="B2">2</xref><xref ref-type="bibr" rid="B3">3</xref><xref ref-type="bibr" rid="B4">4</xref>]. The delivery of high-quality cardiopulmonary resuscitation (CPR), while ensuring chest compression of adequate rate and depth, and allowing full chest recoil, was found to provide a significant survival benefit in previous studies [<xref ref-type="bibr" rid="B5">5</xref><xref ref-type="bibr" rid="B6">6</xref>]. Although the American Heart Association (AHA) and European Resuscitation Council (ERC) guidelines clearly define how to achieve effective CPR in basic life support (BLS) and advanced cardiac life support (ACLS) [<xref ref-type="bibr" rid="B7">7</xref><xref ref-type="bibr" rid="B8">8</xref><xref ref-type="bibr" rid="B9">9</xref>], maintaining good quality of CC in the field has been reported to be challenging [<xref ref-type="bibr" rid="B10">10</xref>].</p>

<p>Therefore, various types of feedback or prompt devices have been introduced to improve the quality of cardiac compression (CC) [<xref ref-type="bibr" rid="B11">11</xref><xref ref-type="bibr" rid="B12">12</xref><xref ref-type="bibr" rid="B13">13</xref><xref ref-type="bibr" rid="B14">14</xref>]. The most updated AHA and ERC guidelines recommend using audiovisual feedback devices during CPR training to improve skill acquisition and retention [<xref ref-type="bibr" rid="B15">15</xref><xref ref-type="bibr" rid="B16">16</xref>]. In the review of interventions to improve CPR quality, not only audiovisual feedback devices but also mobile phones with BLS software or applications also showed favorable outcomes in improving the quality of bystander CPR in the prehospital setting [<xref ref-type="bibr" rid="B17">17</xref>]. However, these feedback devices are not as effective as expected in crowded and noisy environments, especially in the hospital setting [<xref ref-type="bibr" rid="B18">18</xref>].</p>

<p>Recently, a feedback system using a smartwatch was introduced as a substitute [<xref ref-type="bibr" rid="B19">19</xref><xref ref-type="bibr" rid="B20">20</xref><xref ref-type="bibr" rid="B21">21</xref>]. Haptic feedback using vibrations from a smartwatch can be more effectively applicable than an audiovisual feedback device in the crowded and noisy environments of emergency settings [<xref ref-type="bibr" rid="B22">22</xref>]. In addition, smartwatches can give feedback according to each rescuer and can be worn without disturbing the rescuer's movement.</p>

<p>Although previous studies have demonstrated smartwatches as potential feedback devices for high-quality CC, the outcomes are diverse in terms of CPR quality [<xref ref-type="bibr" rid="B19">19</xref><xref ref-type="bibr" rid="B20">20</xref><xref ref-type="bibr" rid="B21">21</xref>]. Furthermore, most of these studies have used either visual or audiovisual feedback in laypersons or medical students. Our study aimed to test the effect of haptic feedback using a smartwatch in the delivery of CC by professional healthcare providers. We hypothesized that haptic feedback using a smartwatch would increase the quality of CPR performed by professional healthcare providers.</p>
</sec>

<sec sec-type="methods">
<title>II. Methods</title>

<sec>
<title>1. Study Design</title>
<p>We designed a prospective, randomized, controlled, standardized simulation study to determine the effect of a haptic device on CPR quality. The study was approved by the Institutional Review Board of Samsung Medical Center (No. 2018-12-014-001).</p>
</sec>

<sec>
<title>2. Study Participants</title>
<p>Twenty participants were recruited from the study site by posting recruitment information on the employee notice board. Medical professionals who were either BLS or ACLS certified were eligible for study participation. Informed consent was obtained from all the participants.</p>
</sec>

<sec>
<title>3. Study Protocol</title>
<p>The study was performed with a case-crossover design. On the first visit, the participants were randomly assigned into two groups. The first group performed 2 minutes of continuous CC according to the AHA guidelines with the device (the haptic-first group), and the other group was asked to perform without it (the control-first group). Then, the participants were summoned 1 week later (washout period) to perform CPR as the other group. The study process is illustrated in <xref ref-type="fig" rid="F1">Figure 1</xref>.</p>
</sec>

<sec>
<title>4. Description of the Devices</title>
<p>An Ambu Man Basic CPR Manikin was used to collect and store the performance data. The R package &#x2018;peakPick&#x2019; was used to detect peaks and intervals of CCs from the performance data. The haptic device used was a Galaxy Gear S3 frontier smartwatch (Samsung Electronics Inc., Seoul, Korea) with a metronome application (Galaxy Store app Wearable Metronome), which is worn at the wrist to provide information with constant metronome-like vibrations at the rate of 110/min (<xref ref-type="fig" rid="F2">Figure 2</xref>). Only the haptic feedback system of the metronome application was used in this study.</p>
</sec>

<sec>
<title>5. Measures</title>
<p>We collected information on demographic factors, such as age, sex, job, and CPR experience. As outcome measures, the total number of CCs was counted during 2 minutes. The participants were assigned to two groups according to their CC performance. The 10 best performing participants without the haptic feedback device were defined as the good performance group (GPG), and the next 10 best performing participants were defined as the bad performance group (BPG). Thirty-second intervals in a sequence were demarcated as quartiles 1Q, 2Q, 3Q, and 4Q.</p>
</sec>

<sec>
<title>6. Primary Outcome</title>
<p>The primary outcome was the percentage of the number of adequate rate of CC, defined as 100&#x2013;120/min, with the haptic feedback device. The total number of CCs was counted during 2 minutes, and between 0.5 and 0.6 seconds for one compression was defined as an adequate duration.</p>
</sec>

<sec>
<title>7. Secondary Outcome</title>
<p>The secondary outcome was a comparative rate of CC and adequate duration between the good and bad performance groups; data obtained at each 30-second quartile were analyzed.</p>
</sec>

<sec>
<title>8. Data Analyses</title>
<p>We described the demographic factors and CPR experience in the two groups. The data were analyzed with the Mann&#x2013; Whitney <italic>U</italic> test, &#x03C7;<sup>2</sup> test, Fisher exact test, or generalized estimation equation (GEE) as appropriate. <italic>p</italic>-values of &#x003C;0.05 were considered statistically significant. The statistical analysis was executed using SAS version 9.4 (SAS Institute, Cary, NC, USA) and R3.5.1. The odds ratio and 95% confidence intervals (CIs) were estimated from the GEE models that included the main variable and covariates. The primary analysis used to examine the haptic device was the main variant of interest. We examined the following four models: model 1, unadjusted analysis between the haptic and control groups; model 2, adjusted analysis between the haptic and control groups; model 3, adjusted as model 2 with additional adjustment for the quality group; and model 4, adjusted as model 3 with additional adjustment for the 4 quartiles.</p>
</sec>

</sec>

<sec sec-type="results">
<title>III. Results</title>
<p>Twenty participants were recruited and then randomized into two groups. Two of the participants were paramedics, 8 were nurses, and 10 were physicians. None of the participants was excluded. The demographic characteristics of the participants are summarized in <xref ref-type="table" rid="T1">Table 1</xref>. No significant differences were found between the two groups regarding sex, age, and experiences.</p>

<sec>
<title>1. Primary Outcome</title>
<p>The CC performance results are shown in <xref ref-type="table" rid="T2">Table 2</xref>. The mean interval of the haptic and non-haptic feedback-assisted CCs did not show a significant difference, and the adequate chest compression number did not show a significant difference. <xref ref-type="fig" rid="F3">Figure 3</xref> shows a comparison of the timeline graphs of CCs for each participant and the fitted values of the haptic and non-haptic feedback-assisted CCs. The timeline graphs of CCs for each participant were more consistent with the haptic feedback device.</p>

<p>In the subgroup analysis between the good and bad performance groups, the mean intervals of the haptic and nonhaptic feedback-assisted CCs&#x2014;good: haptic feedback-assisted (0.57&#x2013;0.06) vs. non-haptic feedback-assisted (0.54&#x2013;0.03), <italic>p</italic> &#x003C; 0.001; bad: haptic feedback-assisted, (0.57&#x2013;0.07) vs. non-haptic feedback assisted (0.58&#x2013;0.18), <italic>p</italic> = 0.005&#x2014;and the adequate chest compression number showed significant differences&#x2014;good: haptic feedback assisted (1,597/75.1%) vs. non-haptic feedback assisted (1,951/92.2%), <italic>p</italic> &#x003C; 0.001; bad: haptic feedback assisted (1,341/63.5%) vs. non-haptic feedback assisted (523/25.4%), <italic>p</italic> &#x003C; 0.001. <xref ref-type="fig" rid="F4">Figure 4</xref> shows a comparison of the timeline graphs of CCs for each participant and the fitted values of the good and bad performance groups between the haptic and non-haptic feedback-assisted CCs. The fitted values in the bad performance group were more consistent with the haptic feedback device.</p>

<p>In another subgroup analysis, the haptic feedback-assisted CCs showed a generally higher quality from 1Q to 4Q, but no statistically significant difference was found in 4Q, unlike in 1Q and 2Q (<xref ref-type="table" rid="T2">Table 2</xref>). The mean CC interval and adequate chest compression numbers with and without haptic assistance were the following, respectively: in 1Q, 0.56&#x2013;0.05 and 0.55&#x2013;0.09 (<italic>p</italic> = 0.016), and 834/76% and 746/67.2% (<italic>p</italic> &#x003C; 0.001); in 2Q, 0.57&#x2013;0.06 and 0.55&#x2013;0.08 (<italic>p</italic> &#x003C; 0.001), and 749/70.7% and 635/58.6% (<italic>p</italic> &#x003C; 0.001); in 3Q, 0.57&#x2013;0.08 and 0.56&#x2013;0.15 (<italic>p</italic> = 0.073), and 719/68.5% and 602/56.6% (<italic>p</italic> &#x003C; 0.001); and in 4Q, 0.58&#x2013;0.07 and 0.57&#x2013;0.18 (<italic>p</italic> = 0.849), and 637/61.4% and 575/57.8% (<italic>p</italic> = 0.11).</p>
</sec>

<sec>
<title>2. Secondary Outcome</title>
<p>In the GEE analysis, the unadjusted odds ratio for the haptic feedback device assistance shown in model 1 (<xref ref-type="table" rid="T3">Table 3</xref>) was 1.502 (95% CI, 0.712&#x2013;3.168). Model 2 shows the adjusted odds ratio (AOR) of the quality group and its positive impact. The AORs were 1.71 (95% CI, 0.69&#x2013;4.24) and 6.80 (95% CI, 3.04&#x2013;15.19) for the GPG and BPG, respectively. The AORs for 2Q, 3Q, and 4Q in comparison with 1Q were 0.67 (95% CI, 0.51&#x2013;0.88), 0.59 (0.41&#x2013;0.87), and 0.50 (0.32&#x2013;0.78), respectively. The adjustment for the quality group had a mixed impact on the odds ratio (model 2 vs. model 3) (<xref ref-type="table" rid="T3">Table 3</xref>). When the haptic feedback device was used, the GPG had a negative impact (AOR = 0.25; 95% CI, 0.11&#x2013;0.59), but the BPG had &#x003E;500% increase in the likelihood of CC of adequate duration (AOR = 5.34; 95% CI, 2.16&#x2013;13.18). In the haptic group, the AOR for the GPG in comparison with the BPG was 1.74 (95% CI, 0.56&#x2013;5.41). In the control group, the AOR for the GPG in comparison with the BPG was 36.80 (95% CI, 14.24&#x2013;95.12).</p>

<p>The AORs for 2Q, 3Q, and 4Q in comparison with 1Q were 0.63 (95% CI, 0.47&#x2013;0.85), 0.55 (0.37&#x2013;0.83), and 0.45 (0.28&#x2013;0.73), respectively. Moreover, both the quality and quartile were adjusted in model 4 (<xref ref-type="table" rid="T3">Table 3</xref>). Each quartile had a positive impact when the haptic feedback device was used (1Q: AOR = 1.74, 95% CI, 0.65&#x2013;4.65; 2Q: AOR = 1.96, 95% CI, 0.75&#x2013;5.11; 3Q: AOR = 1.97, 95% CI, 0.77&#x2013;5.01; 4Q: AOR = 1.29, 95% CI, 0.46&#x2013;3.62). Both quality groups showed a lower probability of high-quality CC in the time sequence. In the GPG, the AORs for 2Q, 3Q, and 4Q in comparison with 1Q were 0.71 (95% CI, 0.50&#x2013;1.02), 0.63 (0.6&#x2013;1.12), and 0.43 (0.21&#x2013;0.89), respectively. In the BPG, the AORs for 2Q, 3Q, and 4Q in comparison with 1Q were 0.63 (95% CI, 0.40&#x2013;0.99), 0.56 (0.33&#x2013;0.95), and 0.582 (0.332&#x2013;1.020), respectively.</p>
</sec>

</sec>

<sec sec-type="discussion">
<title>IV. Discussion</title>
<p>This study demonstrated the capability of a smartwatch using a haptic feedback system for CC. No statistically significant difference in the mean CC duration or the proportion of CCs of adequate duration was found between the two groups. However, the haptic feedback device-assisted group showed less standard deviation than the non-haptic feedback device-assisted group. The compression rates without CPR feedback devices were significantly faster than the recommended guidelines [<xref ref-type="bibr" rid="B23">23</xref>]. In our study, a large number of participants in the non-haptic feedback device-assisted group tended to perform CCs at a faster or slower rate than that recommended in the guidelines or the haptic feedback device-assisted group. This result implies that CPR quality could be homogeneously controlled by using a smartwatch with real-time feedback, and that it is not affected by rescuer experience or the quality of performance.</p>

<p>Previous studies did not show a significant difference in the CC rate in novice rescuers [<xref ref-type="bibr" rid="B20">20</xref><xref ref-type="bibr" rid="B24">24</xref>]. Moreover, even for healthcare professionals, the quality of CPR was suboptimal [<xref ref-type="bibr" rid="B25">25</xref>]. In a subgroup analysis, the authors found that the use of the haptic feedback system of a smartwatch significantly increased the proportion of CCs of adequate duration, especially in the poor performance group. To maintain a homogenously good quality of CC during the entire CPR duration, improving the quality CC in the poor performance group might have great clinical importance.</p>

<p>Recently, the use of smartphones as feedback systems to improve the quality of CC has been introduced in several manikin studies [<xref ref-type="bibr" rid="B26">26</xref>]. Rescuers need to grasp the smartphone in the palm when performing CC, which may cause errors in the accelerometers depending on their grasping methods [<xref ref-type="bibr" rid="B26">26</xref>]. On the other hand, rescuers usually wear smartwatches on the wrist; therefore, if they wear smartwatch, they might overcome the limitations of smartphones. Thus, smartwatches might be better performing feedback devices than smartphones in real-world situations.</p>

<p>The intensity of vibration of smartwatches might vary according to the manufacturer and cannot be customized. The individual perception of vibration might also vary [<xref ref-type="bibr" rid="B27">27</xref>]. Excessive intensity of vibration might interfere with CC performance, and a too weak intensity of vibration might not be perceived. Therefore, further investigation in needed to customize and optimize the intensity of vibration via smartwatches for individual users for better CC performance.</p>

<p>This study had several limitations. First, because this was a simulation study on a manikin, the performance of the compressors would be better than that in real-world clinical practice. The authors assumed the feedback from the watch using vibration would be well delivered even in hectic, noisy environments; however, various features in real-world CPR may interfere with compressor performance.</p>

<p>Second, the AHA guideline recommends a 5- to 6-cm depth and full chest recoil in each compression. The participants were familiar with the guidelines but were reminded again of the guidelines before the study. However, the correct hand position, posture, depth, and chest recoil during CCs were not measured in this study.</p>

<p>Third, the participants in this study were recruited from the emergency department of the study site and were all active healthcare professionals with extensive experience in real-world CPR. Therefore, the performance of compressors might be better than that in real-world. Moreover, the baseline characteristic of the participants, such as their height, weight, or athletic performance were not collected, which may have influenced the quality of CPR.</p>

<p>Finally, this study was conducted for healthcare professionals with hands-only CPR with a compression rate range from 100 to 120 per minute. Therefore, the results might not be applicable in the situation of performing compression to ventilation at a 30:2 ratio.</p>

<p>In conclusion, a CPR feedback system using a smartwatch could not improve rescuers CC performance in terms of the ideal range of CC rate in adherence with the CPR guidelines of the ERC or AHA. In subgroup analysis, the initial compression period of one minute after starting CC might be helpful using a smartwatch to increase the percentage of adequate compression.</p>
</sec>

</body>

<back>

<fn-group>
<fn fn-type="conflict">
<label>Conflict of Interest</label>
<p>No potential conflict of interest relevant to this article was reported.</p>
</fn>
</fn-group>

<ref-list>
 <ref id="B1">
    <label>1</label>
    <element-citation publication-type="journal">
      <person-group person-group-type="author">
        <name>
          <surname>Abella</surname>
          <given-names>BS</given-names>
        </name>
        <name>
          <surname>Alvarado</surname>
          <given-names>JP</given-names>
        </name>
        <name>
          <surname>Myklebust</surname>
          <given-names>H</given-names>
        </name>
        <name>
          <surname>Edelson</surname>
          <given-names>DP</given-names>
        </name>
        <name>
          <surname>Barry</surname>
          <given-names>A</given-names>
        </name>
        <name>
          <surname>O'Hearn</surname>
          <given-names>N</given-names>
        </name>
        <etal/>
      </person-group>
      <article-title>Quality of cardiopulmonary resuscitation during in-hospital cardiac arrest</article-title>
      <source>JAMA</source>
      <year>2005</year>
      <volume>293</volume>
      <issue>3</issue>
      <fpage>305</fpage>
      <lpage>310</lpage>
    </element-citation>
  </ref>

  <ref id="B2">
    <label>2</label>
    <element-citation publication-type="journal">
      <person-group person-group-type="author">
        <name>
          <surname>Eisenberg</surname>
          <given-names>MS</given-names>
        </name>
        <name>
          <surname>Mengert</surname>
          <given-names>TJ</given-names>
        </name>
      </person-group>
      <article-title>Cardiac resuscitation</article-title>
      <source>N Engl J Med</source>
      <year>2001</year>
      <volume>344</volume>
      <issue>17</issue>
      <fpage>1304</fpage>
      <lpage>1313</lpage>
    </element-citation>
  </ref>

  <ref id="B3">
    <label>3</label>
    <element-citation publication-type="journal">
      <person-group person-group-type="author">
        <name>
          <surname>Rea</surname>
          <given-names>TD</given-names>
        </name>
        <name>
          <surname>Crouthamel</surname>
          <given-names>M</given-names>
        </name>
        <name>
          <surname>Eisenberg</surname>
          <given-names>MS</given-names>
        </name>
        <name>
          <surname>Becker</surname>
          <given-names>LJ</given-names>
        </name>
        <name>
          <surname>Lima</surname>
          <given-names>AR</given-names>
        </name>
      </person-group>
      <article-title>Temporal patterns in long-term survival after resuscitation from out-of-hospital cardiac arrest</article-title>
      <source>Circulation</source>
      <year>2003</year>
      <volume>108</volume>
      <issue>10</issue>
      <fpage>1196</fpage>
      <lpage>1201</lpage>
    </element-citation>
  </ref>

  <ref id="B4">
    <label>4</label>
    <element-citation publication-type="journal">
      <person-group person-group-type="author">
        <name>
          <surname>Daya</surname>
          <given-names>MR</given-names>
        </name>
        <name>
          <surname>Schmicker</surname>
          <given-names>RH</given-names>
        </name>
        <name>
          <surname>Zive</surname>
          <given-names>DM</given-names>
        </name>
        <name>
          <surname>Rea</surname>
          <given-names>TD</given-names>
        </name>
        <name>
          <surname>Nichol</surname>
          <given-names>G</given-names>
        </name>
        <name>
          <surname>Buick</surname>
          <given-names>JE</given-names>
        </name>
        <etal/>
      </person-group>
      <article-title>Out-of-hospital cardiac arrest survival improving over time: Results from the Resuscitation Outcomes Consortium (ROC)</article-title>
      <source>Resuscitation</source>
      <year>2015</year>
      <volume>91</volume>
      <fpage>108</fpage>
      <lpage>115</lpage>
    </element-citation>
  </ref>

  <ref id="B5">
    <label>5</label>
    <element-citation publication-type="journal">
      <person-group person-group-type="author">
        <name>
          <surname>Feneley</surname>
          <given-names>MP</given-names>
        </name>
        <name>
          <surname>Maier</surname>
          <given-names>GW</given-names>
        </name>
        <name>
          <surname>Kern</surname>
          <given-names>KB</given-names>
        </name>
        <name>
          <surname>Gaynor</surname>
          <given-names>JW</given-names>
        </name>
        <name>
          <surname>Gall</surname>
          <given-names>SA</given-names>
          <suffix>Jr</suffix>
        </name>
        <name>
          <surname>Sanders</surname>
          <given-names>AB</given-names>
        </name>
        <etal/>
      </person-group>
      <article-title>Influence of compression rate on initial success of resuscitation and 24 hour survival after prolonged manual cardiopulmonary resuscitation in dogs</article-title>
      <source>Circulation</source>
      <year>1988</year>
      <volume>77</volume>
      <fpage>240</fpage>
      <lpage>250</lpage>
    </element-citation>
  </ref>

  <ref id="B6">
    <label>6</label>
    <element-citation publication-type="journal">
      <person-group person-group-type="author">
        <name>
          <surname>Meaney</surname>
          <given-names>PA</given-names>
        </name>
        <name>
          <surname>Bobrow</surname>
          <given-names>BJ</given-names>
        </name>
        <name>
          <surname>Mancini</surname>
          <given-names>ME</given-names>
        </name>
        <name>
          <surname>Christenson</surname>
          <given-names>J</given-names>
        </name>
        <name>
          <surname>de Caen</surname>
          <given-names>AR</given-names>
        </name>
        <name>
          <surname>Bhanji</surname>
          <given-names>F</given-names>
        </name>
        <etal/>
      </person-group>
      <article-title>Cardiopulmonary resuscitation quality: [corrected] improving cardiac resuscitation outcomes both inside and outside the hospital: a consensus statement from the American Heart Association</article-title>
      <source>Circulation</source>
      <year>2013</year>
      <volume>128</volume>
      <issue>4</issue>
      <fpage>417</fpage>
      <lpage>435</lpage>
    </element-citation>
  </ref>

  <ref id="B7">
    <label>7</label>
    <element-citation publication-type="journal">
      <person-group person-group-type="author">
        <name>
          <surname>Kleinman</surname>
          <given-names>ME</given-names>
        </name>
        <name>
          <surname>Brennan</surname>
          <given-names>EE</given-names>
        </name>
        <name>
          <surname>Goldberger</surname>
          <given-names>ZD</given-names>
        </name>
        <name>
          <surname>Swor</surname>
          <given-names>RA</given-names>
        </name>
        <name>
          <surname>Terry</surname>
          <given-names>M</given-names>
        </name>
        <name>
          <surname>Bobrow</surname>
          <given-names>BJ</given-names>
        </name>
        <etal/>
      </person-group>
      <article-title>Part 5: Adult basic life support and cardiopulmonary resuscitation quality: 2015 American Heart Association Guidelines Update for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care</article-title>
      <source>Circulation</source>
      <year>2015</year>
      <volume>132</volume>
      <issue>18 Suppl 2</issue>
      <fpage>S414</fpage>
      <lpage>S435</lpage>
    </element-citation>
  </ref>

  <ref id="B8">
    <label>8</label>
    <element-citation publication-type="journal">
      <person-group person-group-type="author">
        <name>
          <surname>Link</surname>
          <given-names>MS</given-names>
        </name>
        <name>
          <surname>Berkow</surname>
          <given-names>LC</given-names>
        </name>
        <name>
          <surname>Kudenchuk</surname>
          <given-names>PJ</given-names>
        </name>
        <name>
          <surname>Halperin</surname>
          <given-names>HR</given-names>
        </name>
        <name>
          <surname>Hess</surname>
          <given-names>EP</given-names>
        </name>
        <name>
          <surname>Moitra</surname>
          <given-names>VK</given-names>
        </name>
        <etal/>
      </person-group>
      <article-title>Part 7: Adult advanced cardiovascular life support: 2015 American Heart Association Guidelines Update for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care</article-title>
      <source>Circulation</source>
      <year>2015</year>
      <volume>132</volume>
      <issue>18 Suppl 2</issue>
      <fpage>S444</fpage>
      <lpage>S464</lpage>
    </element-citation>
  </ref>

  <ref id="B9">
    <label>9</label>
    <element-citation publication-type="journal">
      <person-group person-group-type="author">
        <name>
          <surname>Perkins</surname>
          <given-names>GD</given-names>
        </name>
        <name>
          <surname>Olasveengen</surname>
          <given-names>TM</given-names>
        </name>
        <name>
          <surname>Maconochie</surname>
          <given-names>I</given-names>
        </name>
        <name>
          <surname>Soar</surname>
          <given-names>J</given-names>
        </name>
        <name>
          <surname>Wyllie</surname>
          <given-names>J</given-names>
        </name>
        <name>
          <surname>Greif</surname>
          <given-names>R</given-names>
        </name>
        <etal/>
      </person-group>
      <article-title>European Resuscitation Council Guidelines for Resuscitation: 2017 update</article-title>
      <source>Resuscitation</source>
      <year>2018</year>
      <volume>123</volume>
      <fpage>43</fpage>
      <lpage>50</lpage>
    </element-citation>
  </ref>

  <ref id="B10">
    <label>10</label>
    <element-citation publication-type="journal">
      <person-group person-group-type="author">
        <name>
          <surname>Wik</surname>
          <given-names>L</given-names>
        </name>
        <name>
          <surname>Kramer-Johansen</surname>
          <given-names>J</given-names>
        </name>
        <name>
          <surname>Myklebust</surname>
          <given-names>H</given-names>
        </name>
        <name>
          <surname>Sorebo</surname>
          <given-names>H</given-names>
        </name>
        <name>
          <surname>Svensson</surname>
          <given-names>L</given-names>
        </name>
        <name>
          <surname>Fellows</surname>
          <given-names>B</given-names>
        </name>
        <etal/>
      </person-group>
      <article-title>Quality of cardiopulmonary resuscitation during out-of-hospital cardiac arrest</article-title>
      <source>JAMA</source>
      <year>2005</year>
      <volume>293</volume>
      <issue>3</issue>
      <fpage>299</fpage>
      <lpage>304</lpage>
    </element-citation>
  </ref>

  <ref id="B11">
    <label>11</label>
    <element-citation publication-type="journal">
      <person-group person-group-type="author">
        <name>
          <surname>Kirkbright</surname>
          <given-names>S</given-names>
        </name>
        <name>
          <surname>Finn</surname>
          <given-names>J</given-names>
        </name>
        <name>
          <surname>Tohira</surname>
          <given-names>H</given-names>
        </name>
        <name>
          <surname>Bremner</surname>
          <given-names>A</given-names>
        </name>
        <name>
          <surname>Jacobs</surname>
          <given-names>I</given-names>
        </name>
        <name>
          <surname>Celenza</surname>
          <given-names>A</given-names>
        </name>
      </person-group>
      <article-title>Audiovisual feedback device use by health care professionals during CPR: a systematic review and meta-analysis of randomised and non-randomised trials</article-title>
      <source>Resuscitation</source>
      <year>2014</year>
      <volume>85</volume>
      <issue>4</issue>
      <fpage>460</fpage>
      <lpage>471</lpage>
    </element-citation>
  </ref>

  <ref id="B12">
    <label>12</label>
    <element-citation publication-type="journal">
      <person-group person-group-type="author">
        <name>
          <surname>Martin</surname>
          <given-names>P</given-names>
        </name>
        <name>
          <surname>Theobald</surname>
          <given-names>P</given-names>
        </name>
        <name>
          <surname>Kemp</surname>
          <given-names>A</given-names>
        </name>
        <name>
          <surname>Maguire</surname>
          <given-names>S</given-names>
        </name>
        <name>
          <surname>Maconochie</surname>
          <given-names>I</given-names>
        </name>
        <name>
          <surname>Jones</surname>
          <given-names>M</given-names>
        </name>
      </person-group>
      <article-title>Real-time feedback can improve infant manikin cardiopulmonary resuscitation by up to 79%: a randomised controlled trial</article-title>
      <source>Resuscitation</source>
      <year>2013</year>
      <volume>84</volume>
      <issue>8</issue>
      <fpage>1125</fpage>
      <lpage>1130</lpage>
    </element-citation>
  </ref>

  <ref id="B13">
    <label>13</label>
    <element-citation publication-type="journal">
      <person-group person-group-type="author">
        <name>
          <surname>Oh</surname>
          <given-names>J</given-names>
        </name>
        <name>
          <surname>Song</surname>
          <given-names>Y</given-names>
        </name>
        <name>
          <surname>Kang</surname>
          <given-names>B</given-names>
        </name>
        <name>
          <surname>Kang</surname>
          <given-names>H</given-names>
        </name>
        <name>
          <surname>Lim</surname>
          <given-names>T</given-names>
        </name>
        <name>
          <surname>Suh</surname>
          <given-names>Y</given-names>
        </name>
        <etal/>
      </person-group>
      <article-title>The use of dual accelerometers improves measurement of chest compression depth</article-title>
      <source>Resuscitation</source>
      <year>2012</year>
      <volume>83</volume>
      <issue>4</issue>
      <fpage>500</fpage>
      <lpage>504</lpage>
    </element-citation>
  </ref>

  <ref id="B14">
    <label>14</label>
    <element-citation publication-type="journal">
      <person-group person-group-type="author">
        <name>
          <surname>Yu</surname>
          <given-names>BG</given-names>
        </name>
        <name>
          <surname>Oh</surname>
          <given-names>JH</given-names>
        </name>
        <name>
          <surname>Kim</surname>
          <given-names>Y</given-names>
        </name>
        <name>
          <surname>Kim</surname>
          <given-names>TW</given-names>
        </name>
      </person-group>
      <article-title>Accurate measurement of chest compression depth using impulseradio ultra-wideband sensor on a mattress</article-title>
      <source>PLoS One</source>
      <year>2017</year>
      <volume>12</volume>
      <issue>8</issue>
      <elocation-id>e0183971</elocation-id>
    </element-citation>
  </ref>

  <ref id="B15">
    <label>15</label>
    <element-citation publication-type="journal">
      <person-group person-group-type="author">
        <name>
          <surname>Bhanji</surname>
          <given-names>F</given-names>
        </name>
        <name>
          <surname>Donoghue</surname>
          <given-names>AJ</given-names>
        </name>
        <name>
          <surname>Wolff</surname>
          <given-names>MS</given-names>
        </name>
        <name>
          <surname>Flores</surname>
          <given-names>GE</given-names>
        </name>
        <name>
          <surname>Halamek</surname>
          <given-names>LP</given-names>
        </name>
        <name>
          <surname>Berman</surname>
          <given-names>JM</given-names>
        </name>
        <etal/>
      </person-group>
      <article-title>Part 14: Education: 2015 American Heart Association Guidelines Update for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care</article-title>
      <source>Circulation</source>
      <year>2015</year>
      <volume>132</volume>
      <issue>18 Suppl 2</issue>
      <fpage>S561</fpage>
      <lpage>S573</lpage>
    </element-citation>
  </ref>

  <ref id="B16">
    <label>16</label>
    <element-citation publication-type="journal">
      <person-group person-group-type="author">
        <name>
          <surname>Greif</surname>
          <given-names>R</given-names>
        </name>
        <name>
          <surname>Lockey</surname>
          <given-names>AS</given-names>
        </name>
        <name>
          <surname>Conaghan</surname>
          <given-names>P</given-names>
        </name>
        <name>
          <surname>Lippert</surname>
          <given-names>A</given-names>
        </name>
        <name>
          <surname>De Vries</surname>
          <given-names>W</given-names>
        </name>
        <name>
          <surname>Monsieurs</surname>
          <given-names>KG</given-names>
        </name>
        <etal/>
      </person-group>
      <article-title>European Resuscitation Council Guidelines for Resuscitation 2015: Section 10. Education and implementation of resuscitation</article-title>
      <source>Resuscitation</source>
      <year>2015</year>
      <volume>95</volume>
      <fpage>288</fpage>
      <lpage>301</lpage>
    </element-citation>
  </ref>

  <ref id="B17">
    <label>17</label>
    <element-citation publication-type="journal">
      <person-group person-group-type="author">
        <name>
          <surname>Chen</surname>
          <given-names>KY</given-names>
        </name>
        <name>
          <surname>Ko</surname>
          <given-names>YC</given-names>
        </name>
        <name>
          <surname>Hsieh</surname>
          <given-names>MJ</given-names>
        </name>
        <name>
          <surname>Chiang</surname>
          <given-names>WC</given-names>
        </name>
        <name>
          <surname>Ma</surname>
          <given-names>MH</given-names>
        </name>
      </person-group>
      <article-title>Interventions to improve the quality of bystander cardiopulmonary resuscitation: a systematic review</article-title>
      <source>PLoS One</source>
      <year>2019</year>
      <volume>14</volume>
      <issue>2</issue>
      <elocation-id>e0211792</elocation-id>
    </element-citation>
  </ref>

  <ref id="B18">
    <label>18</label>
    <element-citation publication-type="journal">
      <person-group person-group-type="author">
        <name>
          <surname>Abella</surname>
          <given-names>BS</given-names>
        </name>
        <name>
          <surname>Edelson</surname>
          <given-names>DP</given-names>
        </name>
        <name>
          <surname>Kim</surname>
          <given-names>S</given-names>
        </name>
        <name>
          <surname>Retzer</surname>
          <given-names>E</given-names>
        </name>
        <name>
          <surname>Myklebust</surname>
          <given-names>H</given-names>
        </name>
        <name>
          <surname>Barry</surname>
          <given-names>AM</given-names>
        </name>
        <etal/>
      </person-group>
      <article-title>CPR quality improvement during inhospital cardiac arrest using a real-time audiovisual feedback system</article-title>
      <source>Resuscitation</source>
      <year>2007</year>
      <volume>73</volume>
      <issue>1</issue>
      <fpage>54</fpage>
      <lpage>61</lpage>
    </element-citation>
  </ref>

  <ref id="B19">
    <label>19</label>
    <element-citation publication-type="confproc">
      <person-group person-group-type="author">
        <name>
          <surname>Gruenerbl</surname>
          <given-names>A</given-names>
        </name>
        <name>
          <surname>Pirkl</surname>
          <given-names>G</given-names>
        </name>
        <name>
          <surname>Monger</surname>
          <given-names>E</given-names>
        </name>
        <name>
          <surname>Gobbi</surname>
          <given-names>M</given-names>
        </name>
        <name>
          <surname>Lukowicz</surname>
          <given-names>P</given-names>
        </name>
      </person-group>
      <source>Smart-watch life saver: smart-watch interactive-feedback system for improving bystander CPR</source>
      <conf-name>Proceedings of the 2015 ACM International Symposium on Wearable Computers</conf-name>
      <conf-date>2015 Sep 7-11</conf-date>
      <conf-loc>Osaka, Japan</conf-loc>
      <fpage>19</fpage>
      <lpage>26</lpage>
    </element-citation>
  </ref>

  <ref id="B20">
    <label>20</label>
    <element-citation publication-type="journal">
      <person-group person-group-type="author">
        <name>
          <surname>Ahn</surname>
          <given-names>C</given-names>
        </name>
        <name>
          <surname>Lee</surname>
          <given-names>J</given-names>
        </name>
        <name>
          <surname>Oh</surname>
          <given-names>J</given-names>
        </name>
        <name>
          <surname>Song</surname>
          <given-names>Y</given-names>
        </name>
        <name>
          <surname>Chee</surname>
          <given-names>Y</given-names>
        </name>
        <name>
          <surname>Lim</surname>
          <given-names>TH</given-names>
        </name>
        <etal/>
      </person-group>
      <article-title>Effectiveness of feedback with a smartwatch for highquality chest compressions during adult cardiac arrest: a randomized controlled simulation study</article-title>
      <source>PLoS One</source>
      <year>2017</year>
      <volume>12</volume>
      <issue>4</issue>
      <elocation-id>e0169046</elocation-id>
    </element-citation>
  </ref>

  <ref id="B21">
    <label>21</label>
    <element-citation publication-type="journal">
      <person-group person-group-type="author">
        <name>
          <surname>Song</surname>
          <given-names>Y</given-names>
        </name>
        <name>
          <surname>Chee</surname>
          <given-names>Y</given-names>
        </name>
        <name>
          <surname>Oh</surname>
          <given-names>J</given-names>
        </name>
        <name>
          <surname>Ahn</surname>
          <given-names>C</given-names>
        </name>
        <name>
          <surname>Lim</surname>
          <given-names>TH</given-names>
        </name>
      </person-group>
      <article-title>Smartwatches as chest compression feedback devices: a feasibility study</article-title>
      <source>Resuscitation</source>
      <year>2016</year>
      <volume>103</volume>
      <fpage>20</fpage>
      <lpage>23</lpage>
    </element-citation>
  </ref>

  <ref id="B22">
    <label>22</label>
    <element-citation publication-type="journal">
      <person-group person-group-type="author">
        <name>
          <surname>Reeder</surname>
          <given-names>B</given-names>
        </name>
        <name>
          <surname>David</surname>
          <given-names>A</given-names>
        </name>
      </person-group>
      <article-title>Health at hand: a systematic review of smart watch uses for health and wellness</article-title>
      <source>J Biomed Inform</source>
      <year>2016</year>
      <volume>63</volume>
      <fpage>269</fpage>
      <lpage>276</lpage>
    </element-citation>
  </ref>

  <ref id="B23">
    <label>23</label>
    <element-citation publication-type="journal">
      <person-group person-group-type="author">
        <name>
          <surname>Lu</surname>
          <given-names>TC</given-names>
        </name>
        <name>
          <surname>Chang</surname>
          <given-names>YT</given-names>
        </name>
        <name>
          <surname>Ho</surname>
          <given-names>TW</given-names>
        </name>
        <name>
          <surname>Chen</surname>
          <given-names>Y</given-names>
        </name>
        <name>
          <surname>Lee</surname>
          <given-names>YT</given-names>
        </name>
        <name>
          <surname>Wang</surname>
          <given-names>YS</given-names>
        </name>
        <etal/>
      </person-group>
      <article-title>Using a smartwatch with real-time feedback improves the delivery of high-quality cardiopulmonary resuscitation by healthcare professionals</article-title>
      <source>Resuscitation</source>
      <year>2019</year>
      <volume>140</volume>
      <fpage>16</fpage>
      <lpage>22</lpage>
    </element-citation>
  </ref>

  <ref id="B24">
    <label>24</label>
    <element-citation publication-type="journal">
      <person-group person-group-type="author">
        <name>
          <surname>Lee</surname>
          <given-names>J</given-names>
        </name>
        <name>
          <surname>Song</surname>
          <given-names>Y</given-names>
        </name>
        <name>
          <surname>Oh</surname>
          <given-names>J</given-names>
        </name>
        <name>
          <surname>Chee</surname>
          <given-names>Y</given-names>
        </name>
        <name>
          <surname>Ahn</surname>
          <given-names>C</given-names>
        </name>
        <name>
          <surname>Shin</surname>
          <given-names>H</given-names>
        </name>
        <etal/>
      </person-group>
      <article-title>Smartwatch feedback device for high-quality chest compressions by a single rescuer during infant cardiac arrest: a randomized, controlled simulation study</article-title>
      <source>Eur J Emerg Med</source>
      <year>2019</year>
      <volume>26</volume>
      <issue>4</issue>
      <fpage>266</fpage>
      <lpage>271</lpage>
    </element-citation>
  </ref>

  <ref id="B25">
    <label>25</label>
    <element-citation publication-type="journal">
      <person-group person-group-type="author">
        <name>
          <surname>Abella</surname>
          <given-names>BS</given-names>
        </name>
        <name>
          <surname>Sandbo</surname>
          <given-names>N</given-names>
        </name>
        <name>
          <surname>Vassilatos</surname>
          <given-names>P</given-names>
        </name>
        <name>
          <surname>Alvarado</surname>
          <given-names>JP</given-names>
        </name>
        <name>
          <surname>O'Hearn</surname>
          <given-names>N</given-names>
        </name>
        <name>
          <surname>Wigder</surname>
          <given-names>HN</given-names>
        </name>
        <etal/>
      </person-group>
      <article-title>Chest compression rates during cardiopulmonary resuscitation are suboptimal: a prospective study during in-hospital cardiac arrest</article-title>
      <source>Circulation</source>
      <year>2005</year>
      <volume>111</volume>
      <issue>4</issue>
      <fpage>428</fpage>
      <lpage>434</lpage>
    </element-citation>
  </ref>

  <ref id="B26">
    <label>26</label>
    <element-citation publication-type="journal">
      <person-group person-group-type="author">
        <name>
          <surname>Song</surname>
          <given-names>Y</given-names>
        </name>
        <name>
          <surname>Oh</surname>
          <given-names>J</given-names>
        </name>
        <name>
          <surname>Chee</surname>
          <given-names>Y</given-names>
        </name>
      </person-group>
      <article-title>A new chest compression depth feedback algorithm for high-quality CPR based on smartphone</article-title>
      <source>Telemed J E Health</source>
      <year>2015</year>
      <volume>21</volume>
      <issue>1</issue>
      <fpage>36</fpage>
      <lpage>41</lpage>
    </element-citation>
  </ref>

  <ref id="B27">
    <label>27</label>
    <element-citation publication-type="journal">
      <person-group person-group-type="author">
        <name>
          <surname>Lundstrom</surname>
          <given-names>R</given-names>
        </name>
        <name>
          <surname>Stromberg</surname>
          <given-names>T</given-names>
        </name>
        <name>
          <surname>Lundborg</surname>
          <given-names>G</given-names>
        </name>
      </person-group>
      <article-title>Vibrotactile perception threshold measurements for diagnosis of sensory neuropathy: description of a reference population</article-title>
      <source>Int Arch Occup Environ Health</source>
      <year>1992</year>
      <volume>64</volume>
      <issue>3</issue>
      <fpage>201</fpage>
      <lpage>207</lpage>
    </element-citation>
  </ref>

</ref-list>

</back>

<floats-group>

<fig position="float" id="F1">
<label>Figure 1</label>
<caption>
  <title>CONSORT flow diagram for the study.</title>
</caption>
<graphic xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="hir-25-274-g001"></graphic>
</fig>

<fig position="float" id="F2">
<label>Figure 2</label>
<caption>
  <title>The smartwatch screen displays the rate of 110/min.</title>
</caption>
<graphic xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="hir-25-274-g002"></graphic>
</fig>

<fig position="float" id="F3">
<label>Figure 3</label>
<caption>
<title>Comparison of the timeline graphs of cardiac compressions (CCs) in each participant and the fitted values of the haptic and non-haptic feedback-assisted CCs.</title>
</caption>
<graphic xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="hir-25-274-g003"></graphic>
</fig>

<fig position="float" id="F4">
<label>Figure 4</label>
<caption>
<title>Comparison of the timeline graphs of cardiac compressions (CCs) in each participant and the fitted values of the good and bad performance groups between the haptic and non-haptic feedback-assisted CCs.</title>
</caption>
<graphic xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="hir-25-274-g004"></graphic>
</fig>

<table-wrap position="float" id="T1">
<label>Table 1</label>
<caption>
  <title>Demographic characteristics of the participants</title>
</caption>
<graphic xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="hir-25-274-i001"></graphic>
<table-wrap-foot>
<fn>
  <p>Values are presented as mean &#x00B1; standard deviation or number (%).</p>
  <p>BLS: basic life support, ACLS: advanced cardiovascular life support, EMT: emergency medical technician.</p>
</fn>
</table-wrap-foot>
</table-wrap>

<table-wrap position="float" id="T2">
<label>Table 2</label>
<caption>
  <title>Results for cardiac compression performance</title>
</caption>
<graphic xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="hir-25-274-i002"></graphic>
<table-wrap-foot>
<fn>
  <p>Values are presented as mean &#x00B1; standard deviation or number (%).</p>
<p>Comparative rate of CC and adequate duration between the good and bad performance groups; data obtained at each 30-second quartile were analyzed.</p>
  <p>CC: cardiac compression.</p>
</fn>
</table-wrap-foot>
</table-wrap>

<table-wrap position="float" id="T3">
<label>Table 3</label>
<caption>
  <title>Results from the generalized estimating equation model</title>
</caption>
<graphic xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="hir-25-274-i003"></graphic>
<table-wrap-foot>
<fn>
<p>Model 1, unadjusted analysis between the haptic and control groups; Model 2, adjusted analysis between the haptic and control groups; Model 3, adjusted as Model 2 with additional adjustment for the quality group; and Model 4, adjusted as Model 3 with additional adjustment for the 4 quartiles; OR: odds ratio, CI: confidence interval, CC: cardiac compression, GPG: good performance group, BPG: bad performance group.</p>
</fn>
</table-wrap-foot>
</table-wrap>

</floats-group>

</article>