<article-meta>
<article-id pub-id-type="doi">10.14316/pmp.2018.29.4.123</article-id>
<article-id pub-id-type="publisher-id">pmp-29-123</article-id>
<article-categories>
<subj-group subj-group-type="heading">
<subject>Original Article</subject>
</subj-group>
</article-categories>
<title-group>
<article-title>Guideline on Acceptance Test and Commissioning of High-Precision External Radiation Therapy Equipment</article-title>
</title-group>
<contrib-group>
<contrib contrib-type="author">
<name name-style="western" xml:lang="en"><surname>Kim</surname><given-names>Juhye</given-names></name>
<xref ref-type="aff" rid="aff1-pmp-29-123"><sup>&#x2217;</sup></xref>
</contrib>
<contrib contrib-type="author">
<name name-style="western" xml:lang="en"><surname>Shin</surname><given-names>Dong Oh</given-names></name><xref ref-type="aff" rid="aff2-pmp-29-123"><sup>&#x2020;</sup></xref>
</contrib>
<contrib contrib-type="author">
<name name-style="western" xml:lang="en"><surname>Choi</surname><given-names>Sang Hyoun</given-names></name><xref ref-type="aff" rid="aff3-pmp-29-123"><sup>&#x2021;</sup></xref>
</contrib>
<contrib contrib-type="author">
<name name-style="western" xml:lang="en"><surname>Min</surname><given-names>Soonki</given-names></name><xref ref-type="aff" rid="aff2-pmp-29-123"><sup>&#x2020;</sup></xref>
</contrib>
<contrib contrib-type="author">
<name name-style="western" xml:lang="en"><surname>Kwon</surname><given-names>Nahye</given-names></name><xref ref-type="aff" rid="aff4-pmp-29-123"><sup>&#x00A7;</sup></xref>
</contrib>
<contrib contrib-type="author">
<name name-style="western" xml:lang="en"><surname>Jung</surname><given-names>Unjung</given-names></name>
<xref ref-type="aff" rid="aff1-pmp-29-123"><sup>&#x2217;</sup></xref>
</contrib>
<contrib contrib-type="author">
<name name-style="western" xml:lang="en"><surname>Kim</surname><given-names>Dong Wook</given-names></name>
<xref ref-type="aff" rid="aff5-pmp-29-123"><sup>&#x0399;&#x0399;</sup></xref>
<xref ref-type="corresp" rid="c1-pmp-29-123"/>
</contrib>
<aff id="aff1-pmp-29-123"><label>&#x2217;</label>Research Institute of Clinical Medicine, Kyung Hee University Hospital at Gangdong</aff>
<aff id="aff2-pmp-29-123"><label>&#x2020;</label>Department of Radiation Oncology, Kyung Hee University Hospital</aff>
<aff id="aff3-pmp-29-123"><label>&#x2021;</label>Division of Medical Radiation Equipment, Korea Institute of Radiological and Medical Sciences, Seoul</aff>
<aff id="aff4-pmp-29-123"><label>&#x00A7;</label>Department of Nuclear Engineering, Kyung Hee University, Yongin</aff>
<aff id="aff5-pmp-29-123"><label>&#x0399;&#x0399;</label>Department of Radiation Oncology, Kyung Hee University Hospital at Gangdong, Seoul, <country>Korea</country></aff>
</contrib-group>
<author-notes>
<corresp id="c1-pmp-29-123"><bold>Corresponding author</bold> Dong Wook Kim (<email>joocheck@gmail.com</email>) Tel: 82-2-440-7398 Fax: 82-2-440-7393</corresp></author-notes>
<pub-date pub-type="ppub"><month>01</month><year>2018</year></pub-date>
<pub-date pub-type="epub"><day>19</day><month>01</month><year>2018</year></pub-date>
<volume>29</volume><issue>4</issue><fpage>123</fpage>
<lpage>136</lpage>
<history>
<date date-type="received"><day>20</day><month>11</month><year>2018</year></date>
<date date-type="rev-recd"><day>13</day><month>12</month><year>2018</year></date>
<date date-type="accepted"><day>14</day><month>12</month><year>2018</year></date>
</history>
<permissions>
<copyright-statement>Copyright &#x00A9; 2018 Korean Society of Medical Physics</copyright-statement>
<copyright-year>2018</copyright-year>
<license><license-p>This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (<ext-link ext-link-type="uri" xlink:href="http://creativecommons.org/licenses/by-nc/3.0">http://creativecommons.org/licenses/by-nc/3.0</ext-link>) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.</license-p></license>
</permissions>
<abstract xml:lang="en">
<p>The complex dose distribution and dose transfer characteristics of intensity-modulated radiotherapy increase the importance of precise beam data measurement and review in the acceptance inspection and preparation stages. In this study, we propose a process map for the introduction and installation of high-precision radiotherapy devices and present items and guidelines for risk management at the acceptance test procedure (ATP) and commissioning stages. Based on the ATP of the Varian and Elekta linear accelerators, the ATP items were checked step by step and compared with the quality assurance (QA) test items of the AAPM TG-142 described for the medical accelerator QA. Based on the commissioning procedure, dose quality control protocol, and mechanical quality control protocol presented at international conferences, step-by-step check items and commissioning guidelines were derived. The risk management items at each stage were (1) 21 ionization chamber performance test items and 9 electrometer, cable, and connector inspection items related to the dosimetry system; (2) 34 mechanical and dose-checking items during ATP, 22 multileaf collimator (MLC) items, and 36 imaging system items; and (3) 28 items in the measurement preparation stage and 32 items in the measurement stage after commissioning. Because the items presented in these guidelines are limited in terms of special treatment, items and practitioners can be modified to reflect the clinical needs of the institution. During the system installation, it is recommended that at least two clinically qualified medical physicists (CQMP) perform a double check in compliance with the two-person rule. We expect that this result will be useful as a radiation safety management tool that can prevent radiation accidents at each stage during the introduction of radiotherapy and the system installation process.</p>
</abstract>
<kwd-group xml:lang="en">
<kwd>External radiation therapy equipment</kwd>
<kwd>Acceptance test</kwd>
<kwd>Commissioning</kwd>
<kwd>IMRT</kwd>
<kwd>Risk management</kwd>
</kwd-group>
</article-meta>
</front>
<back><ref-list xml:lang="en"><title>REFERENCES</title><ref id="b1-pmp-29-123"><element-citation publication-type="journal"><label>1.</label><person-group person-group-type="author"><name><surname>Gary</surname> <given-names>A</given-names></name> <name><surname>Ezzell</surname></name> <name><surname>Calvin</surname> <given-names>James M.</given-names></name> <name><surname>Low</surname> <given-names>Daniel</given-names></name> <name><surname>Rosen</surname> <given-names>Isaac</given-names></name> <name><surname>Sharpe</surname> <given-names>Michael B.</given-names></name> <name><surname>Xia</surname> <given-names>Ping</given-names></name> <etal/></person-group> <article-title>Guidance document on delivery, treatment planning, and clinical implementation of IMRT: Report of the IMRT subcommittee of the AAPM radiation therapy committee</article-title> <source>Med. Phys.</source> <year>2003</year><volume>30</volume><issue>8</issue> <fpage>2089</fpage><lpage>2115</lpage></element-citation></ref>
<ref id="b2-pmp-29-123"><element-citation publication-type="other"><label>2.</label><article-title>IAEA Human Health Series No.25: Roles and Responsibilities, and Education and Training Requirements for Clinically Qualified Medical Physicists</article-title> <source>International Atomic Energy Agency.</source> <year>2013</year></element-citation></ref>
<ref id="b3-pmp-29-123"><element-citation publication-type="other"><label>3.</label><collab>IAEA TRS-398 report: Absorbed Dose Determination in External Beam Radiotherapy: An International Code of Practice for Dosimetry based on Standards of Absorbed Dose to Water</collab> <source>International Atomic Energy Agency.</source> <year>2006</year></element-citation></ref>
<ref id="b4-pmp-29-123"><element-citation publication-type="journal"><label>4.</label><person-group person-group-type="author"><name><surname>Almond</surname> <given-names>P. R.</given-names></name> <name><surname>Biggs</surname> <given-names>P. J.</given-names></name> <name><surname>Coursey</surname> <given-names>B. M.</given-names></name> <name><surname>Hanson</surname> <given-names>W. F.</given-names></name> <name><surname>Saiful Huq</surname> <given-names>M.</given-names></name> <name><surname>Nath</surname> <given-names>Ravinder</given-names></name> <etal/></person-group> <article-title>AAPM&#x0027;s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams</article-title> <source>Med. Phys.</source> <year>1999</year><volume>26</volume><issue>9</issue> <fpage>1847</fpage><lpage>1870</lpage></element-citation></ref>
<ref id="b5-pmp-29-123"><element-citation publication-type="journal"><label>5.</label><person-group person-group-type="author"><name><surname>Gerald</surname> <given-names>J</given-names></name> <name><surname>Kutche</surname></name> <name><surname>Coia</surname> <given-names>Lawrene</given-names></name> <name><surname>Gillin</surname> <given-names>Michael</given-names></name> <name><surname>Hansom</surname> <given-names>Wiliam F.</given-names></name> <name><surname>Leibel</surname> <given-names>Steven</given-names></name> <name><surname>Morton</surname> <given-names>Robert J.</given-names></name> <etal/></person-group> <article-title>Comprehensive QA for radiation oncology y: Report of AAPM Radiation Therapy Committee Task Group 40</article-title> <source>Med. Phys.</source> <year>1994</year><volume>21</volume><issue>4</issue> <fpage>581</fpage><lpage>618</lpage></element-citation></ref>
<ref id="b6-pmp-29-123"><element-citation publication-type="journal"><label>6.</label><person-group person-group-type="author"><name><surname>Eric</surname> <given-names>E</given-names></name> <name><surname>Klein</surname></name> <name><surname>Hanley</surname> <given-names>Joseph</given-names></name> <name><surname>Bayouth</surname> <given-names>John</given-names></name> <name><surname>Yin</surname> <given-names>Fang-Fang</given-names></name> <name><surname>Simon</surname> <given-names>Wiliam</given-names></name> <name><surname>Dresser</surname> <given-names>Sen</given-names></name> <etal/></person-group> <article-title>Task Group 142 report: Quality assurance of medical accelerators</article-title> <source>Med. Phys.</source> <year>2009</year><volume>36</volume><issue>9</issue> <fpage>4197</fpage><lpage>4212</lpage></element-citation></ref>
<ref id="b7-pmp-29-123"><element-citation publication-type="journal"><label>7.</label><person-group person-group-type="author"><name><surname>Indra</surname> <given-names>J</given-names></name> <name><surname>Das</surname></name> <name><surname>Cheng</surname> <given-names>Chee-Wai</given-names></name> <name><surname>Watts</surname> <given-names>Ronald J.</given-names></name> <name><surname>Ahnesj&#x00F6;</surname> <given-names>Anders</given-names></name> <name><surname>Gibbons</surname> <given-names>John</given-names></name> <name><surname>Allen Li</surname> <given-names>X.</given-names></name> <etal/></person-group> <article-title>Accelerator beam data commissioning equipment and procedures: Report of the TG-106 of the Therapy Physics Committee of the AAPM</article-title> <source>Med. Phys.</source> <year>2008</year><volume>35</volume><issue>9</issue> <fpage>4186</fpage><lpage>4215</lpage></element-citation></ref>
<ref id="b8-pmp-29-123"><element-citation publication-type="journal"><label>8.</label><person-group person-group-type="author"><name><surname>Gary</surname> <given-names>A</given-names></name> <name><surname>Ezzell</surname></name> <name><surname>Burmeister</surname> <given-names>Jay W.</given-names></name> <name><surname>Dogan</surname> <given-names>Nesrin</given-names></name> <name><surname>LoSasso</surname> <given-names>Thomas J.</given-names></name> <name><surname>Mechalakos</surname> <given-names>James G.</given-names></name> <name><surname>Mihailidis</surname> <given-names>Dimitiris</given-names></name> <etal/></person-group> <article-title>IMRT commissioning: Multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119</article-title> <source>Med. Phys.</source> <year>2009</year><volume>36</volume><issue>11</issue> <fpage>5359</fpage><lpage>5373</lpage></element-citation></ref>
<ref id="b9-pmp-29-123"><element-citation publication-type="other"><label>9.</label><source>ESTRO booklet no.9: Guidelines for the verification of IMRT</source> <year>2008</year></element-citation></ref>
<ref id="b10-pmp-29-123"><element-citation publication-type="journal"><label>10.</label><person-group person-group-type="author"><name><surname>Daniel</surname> <given-names>A</given-names></name></person-group> <article-title>Low, Jean M. Moran, James F. Dempsey, Lei Dong, Mark Oldham, Dosimetry tools and techniques for IMRT</article-title> <source>Med. Phys.</source> <year>2001</year><volume>38</volume><issue>3</issue> <fpage>1313</fpage><lpage>1338</lpage></element-citation></ref>
<ref id="b11-pmp-29-123"><element-citation publication-type="journal"><label>11.</label><person-group person-group-type="author"><name><surname>Oh</surname> <given-names>Yoonjin</given-names></name> <name><surname>Shin</surname> <given-names>Dong Oh</given-names></name> <name><surname>Kim</surname> <given-names>Juhye</given-names></name> <name><surname>Kwon</surname> <given-names>Nahye</given-names></name> <name><surname>Lee</surname> <given-names>Soon Sung</given-names></name> <name><surname>Choi</surname> <given-names>Sang Hyoun</given-names></name> <etal/></person-group> <article-title>Proposal on Guideline for Quality Assurance of Radiation Treatment Planning System</article-title> <source>Progress in Medical Physics.</source> <year>2017</year><volume>28</volume><issue>4</issue> <fpage>197</fpage><lpage>206</lpage></element-citation></ref>
<ref id="b12-pmp-29-123"><element-citation publication-type="other"><label>12.</label><collab>L</collab> <article-title>J. Humphries and J</article-title> <source>A. Purdy, In Advances in Radiation Oncology Physics Dosimetry, Treatment Planning, and Brachytherapy.</source> <year>1992</year> <comment>AAPM Monograph Vol. 19, edited by J. A. Purdy AAPM.</comment></element-citation></ref>
<ref id="b13-pmp-29-123"><element-citation publication-type="journal"><label>13.</label><person-group person-group-type="author"><name><surname>Per</surname> <given-names>H</given-names></name> <name><surname>Halvorsen</surname></name> <name><surname>Das</surname> <given-names>Indra J.</given-names></name> <name><surname>Fraser</surname> <given-names>Martin</given-names></name> <name><surname>Jay Freedman</surname> <given-names>D.</given-names></name> <name><surname>Robert E. Rice</surname> <given-names>III</given-names></name> <name><surname>lbbott</surname> <given-names>Geoffrey S.</given-names></name> <etal/></person-group> <article-title>AAPM Task Group 103 report on peer review in clinical radiation oncology physics</article-title> <source>Journal of Applied Clinical Medical Physics.</source> <year>2005</year><volume>6</volume><issue>4</issue> <fpage>50</fpage><lpage>64</lpage></element-citation></ref></ref-list>
<sec sec-type="display-objects">
<title>Figures and Tables</title>
<fig id="f1-pmp-29-123" position="float">
<label>Fig. 1</label>
<caption xml:lang="en"><p>Ionization chamber performance test.</p></caption>
<graphic xlink:href="pmp-29-123f1.tif"/>
</fig>
<fig id="f2-pmp-29-123" position="float">
<label>Fig. 2</label>
<caption xml:lang="en"><p>The risk management items for electrometer, cable and connector check.</p></caption>
<graphic xlink:href="pmp-29-123f2.tif"/>
</fig>
<fig id="f3-pmp-29-123" position="float">
<label>Fig. 3</label>
<caption xml:lang="en"><p>The risk management items for acceptance test of external radiation therapy equipment: Mechanical and dosimetry test.</p></caption>
<graphic xlink:href="pmp-29-123f3.tif"/>
</fig>
<fig id="f4-pmp-29-123" position="float">
<label>Fig. 4</label>
<caption xml:lang="en"><p>The risk management items for acceptance test of external radiation therapy equipment: MLC.</p></caption>
<graphic xlink:href="pmp-29-123f4.tif"/>
</fig>
<fig id="f5-pmp-29-123" position="float">
<label>Fig. 5</label>
<caption xml:lang="en"><p>The risk management items for acceptance test of external radiation therapy equipment: Imaging system.</p></caption>
<graphic xlink:href="pmp-29-123f5.tif"/>
</fig>
<fig id="f6-pmp-29-123" position="float">
<label>Fig. 6</label>
<caption xml:lang="en"><p>The risk management items for commissioning of external radiation therapy equipment: preparation of beam measurement.</p></caption>
<graphic xlink:href="pmp-29-123f6.tif"/>
</fig>
<fig id="f7-pmp-29-123" position="float">
<label>Fig. 7</label>
<caption xml:lang="en"><p>The risk management items for commissioning of external radiation therapy equipment: beam measurement.</p></caption>
<graphic xlink:href="pmp-29-123f7.tif"/>
</fig>
<table-wrap id="t1-pmp-29-123" position="float">
<label>Table 1.</label>
<caption xml:lang="en"><p>The derived risk management items were compared for the correlation with TG-142 quality assurance items.</p></caption>
<table frame="hsides" rules="all">
<thead>
<tr>
<th valign="top" align="center">Step</th>
<th valign="top" align="center"></th>
<th valign="top" align="center" colspan="2">Mechanical and Dosimetry check</th>
<th valign="top" align="center"></th>
<th valign="top" align="center"></th>
</tr>
<tr>
<th valign="top" align="center">Sub-step</th>
<th valign="top" align="center">Risk management items</th>
<th valign="top" align="center">TG-142 Quality assurance items</th>
<th valign="top" align="center">Period</th>
<th valign="top" align="center">IMRT Tolerance</th>
<th valign="top" align="center">Remark</th>
</tr>
</thead>
<tbody>
<tr>
<td valign="top" align="left">Mechanical test</td>
<td valign="top" align="left">Mechanical isocenter variation (with collimator, gantry, couch rotation)</td>
<td valign="top" align="left">Gantry/collimator angle indicators (digital only)</td>
<td valign="top" align="left">Monthly</td>
<td valign="top" align="left">1&#x00B0;</td>
<td valign="top" align="left">Cardinal angles</td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left">Treatment couch position indicator</td>
<td valign="top" align="left">Monthly</td>
<td valign="top" align="left">2 mm/1&#x00B0;</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left">Collimator rotation isocenter</td>
<td valign="top" align="left">Annual</td>
<td valign="top" align="left">&#x00B1;1 mm</td>
<td valign="top" align="left">Change from baseline</td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left">Gantry rotation isocenter</td>
<td valign="top" align="left">Annual</td>
<td valign="top" align="left">&#x00B1;1 mm</td>
<td valign="top" align="left">Change from baseline</td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left">Couch rotation isocenter</td>
<td valign="top" align="left">Annual</td>
<td valign="top" align="left">&#x00B1;1 mm</td>
<td valign="top" align="left">Change from baseline</td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left">Cross-hair check</td>
<td valign="top" align="left">Cross-hair centering (walkout)</td>
<td valign="top" align="left">Monthly</td>
<td valign="top" align="left">1 mm</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left">Digital display indicator calibration</td>
<td valign="top" align="left">Collimator size indicator</td>
<td valign="top" align="left">Daily</td>
<td valign="top" align="left">2 mm</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left">&#x00A0;(Jaw position, gantry, collimator rotation)</td>
<td valign="top" align="left">Gantry/collimator angle indicators (digital only)</td>
<td valign="top" align="left">Monthly</td>
<td valign="top" align="left">1&#x00B0;</td>
<td valign="top" align="left">Cardinal angles</td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left">Jaw position indicators (symmetry)</td>
<td valign="top" align="left">Monthly</td>
<td valign="top" align="left">2 mm</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left">Jaw position indicators (asymmetry)</td>
<td valign="top" align="left">Monthly</td>
<td valign="top" align="left">1 mm</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left">Couch movement (rotation,</td>
<td valign="top" align="left">Treatment couch position indicator</td>
<td valign="top" align="left">Monthly</td>
<td valign="top" align="left">2 mm/1&#x00B0;</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left">longitudinal, lateral, vertical)</td>
<td valign="top" align="left">Table top sag</td>
<td valign="top" align="left">Annual</td>
<td valign="top" align="left">2 mm</td>
<td valign="top" align="left">Change from baseline</td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left">Table angle</td>
<td valign="top" align="left">Annual</td>
<td valign="top" align="left">1&#x00B0;</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left">Table travel maximum range movement</td>
<td valign="top" align="left">Annual</td>
<td valign="top" align="left">&#x00B1;2 mm</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left">in all directions</td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left">Optical distance indicator (ODI)</td>
<td valign="top" align="left">Distance indicator (ODI) at isocenter</td>
<td valign="top" align="left">Daily</td>
<td valign="top" align="left">2 mm</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left">verification</td>
<td valign="top" align="left">Distance check device for lasers</td>
<td valign="top" align="left">Monthly</td>
<td valign="top" align="left">1 mm</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left">compared with front pointer</td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left">Radiation isocenter</td>
<td valign="top" align="left">Spoke shot (Gantry rotation)</td>
<td valign="top" align="left">Coincidence of radiation and mechanical</td>
<td valign="top" align="left">l Annual</td>
<td valign="top" align="left">&#x00B1;2 mm</td>
<td valign="top" align="left">Change from baseline</td>
</tr>
<tr>
<td valign="top" align="left">check</td>
<td valign="top" align="left"></td>
<td valign="top" align="left">isocenter</td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left">Winston-Lutz test (with gantry/</td>
<td valign="top" align="left">Collimator rotation isocenter</td>
<td valign="top" align="left">Annual</td>
<td valign="top" align="left">&#x00B1;1 mm</td>
<td valign="top" align="left">Change from baseline</td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left">collimator/couch rotation)</td>
<td valign="top" align="left">Gantry rotation isocenter</td>
<td valign="top" align="left">Annual</td>
<td valign="top" align="left">&#x00B1;1 mm</td>
<td valign="top" align="left">Change from baseline</td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left">Couch rotation isocenter</td>
<td valign="top" align="left">Annual</td>
<td valign="top" align="left">&#x00B1;1 mm</td>
<td valign="top" align="left">Change from baseline</td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left">Coincidence of light field and X-ray</td>
<td valign="top" align="left">Light/radiation field coincidence</td>
<td valign="top" align="left">Monthly</td>
<td valign="top" align="left">2 mm or 1%</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left">field</td>
<td valign="top" align="left">(symmetry)</td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left">Light/radiation field coincidence</td>
<td valign="top" align="left">Monthly</td>
<td valign="top" align="left">1 mm or 1%</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left">(asymmetry)</td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left">Laser guard collision</td>
<td valign="top" align="left">Protection zone area verification</td>
<td valign="top" align="left">Laser guard-interlock test</td>
<td valign="top" align="left">Monthly</td>
<td valign="top" align="left">Functional</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left">protection system</td>
<td valign="top" align="left">Protection zone tilt verification</td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left">Motion stop function verification</td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left">Power key switch and override function verification</td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left">Beam performance</td>
<td valign="top" align="left">Photon PDD</td>
<td valign="top" align="left">X-ray beam quality (PDD<sub>10</sub> or TMR<sub>20,10</sub>)</td>
<td valign="top" align="left">Annual</td>
<td valign="top" align="left">&#x00B1;1%</td>
<td valign="top" align="left">Change from baseline</td>
</tr>
<tr>
<td valign="top" align="left">test</td>
<td valign="top" align="left">Photon field flatness/symmetry</td>
<td valign="top" align="left">Photon beam profile constancy</td>
<td valign="top" align="left">Monthly</td>
<td valign="top" align="left">1%</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left">X-ray flatness change from baseline</td>
<td valign="top" align="left">Annual</td>
<td valign="top" align="left">1%</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left">X-ray symmetry change from baseline</td>
<td valign="top" align="left">Annual</td>
<td valign="top" align="left">&#x00B1;1%</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left">Electron PDI</td>
<td valign="top" align="left">Electron energy constancy</td>
<td valign="top" align="left">Monthly</td>
<td valign="top" align="left">2%/2 mm</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left">Electron beam quality (R<sub>50</sub>)</td>
<td valign="top" align="left">Annual</td>
<td valign="top" align="left">&#x00B1;1 mm</td>
<td valign="top" align="left">Change from baseline</td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left">Electron field flatness/symmetry</td>
<td valign="top" align="left">Electron beam profile constancy</td>
<td valign="top" align="left">Monthly</td>
<td valign="top" align="left">1%</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left">Electron flatness change from baseline</td>
<td valign="top" align="left">Annual</td>
<td valign="top" align="left">1%</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left">Electron symmetry change from baseline</td>
<td valign="top" align="left">Annual</td>
<td valign="top" align="left">&#x00B1;1%</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left">Symmetry interlock check</td>
<td valign="top" align="left">Follow manufacturer&#x0027;s test procedure</td>
<td valign="top" align="left">Annual</td>
<td valign="top" align="left">Functional</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left">Beamstopper interlocked angles</td>
<td valign="top" align="left">Follow manufacturer&#x0027;s test procedure</td>
<td valign="top" align="left">Annual</td>
<td valign="top" align="left">Functional</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left">Dosimetry</td>
<td valign="top" align="left">Dose reproducibility</td>
<td valign="top" align="left">X-ray output constancy</td>
<td valign="top" align="left">Daily</td>
<td valign="top" align="left">3%</td>
<td valign="top" align="left">All energies</td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left">Monthly</td>
<td valign="top" align="left">2%</td>
<td valign="top" align="left">All energies</td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left">Electron output constancy</td>
<td valign="top" align="left">Daily</td>
<td valign="top" align="left">3%</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left">Monthly</td>
<td valign="top" align="left">2%</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left">Backup monitor chamber constancy</td>
<td valign="top" align="left">Monthly</td>
<td valign="top" align="left">2%</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left">X-ray/electron output calibration (TG-51)</td>
<td valign="top" align="left">) Annual</td>
<td valign="top" align="left">&#x00B1;1%</td>
<td valign="top" align="left">Absolute dosimetry</td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left">Spot check of field size dependent output</td>
<td valign="top" align="left">Annual</td>
<td valign="top" align="left">2%</td>
<td valign="top" align="left">Two or more field size</td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left">factors for X-ray</td>
<td valign="top" align="left"></td>
<td valign="top" align="left">&#x003C;4&#x00D7;4 cm<sup>2</sup></td>
<td valign="top" align="left">check</td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left">1% &#x2265;4&#x00D7;4 cm<sup>2</sup></td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left">Output factors for electron applicators</td>
<td valign="top" align="left">Annual</td>
<td valign="top" align="left">&#x00B1;2%</td>
<td valign="top" align="left">- Change from baseline</td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left">- Spot check of on</td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left">applicator/energy</td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left">Dose linearity with MU</td>
<td valign="top" align="left">X-ray monitor unit linearity output</td>
<td valign="top" align="left">Annual</td>
<td valign="top" align="left">&#x00B1;2%</td>
<td valign="top" align="left">&#x00B1;5% (2-4 MU)</td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left">constancy</td>
<td valign="top" align="left"></td>
<td valign="top" align="left">(&#x2265;5 MU)</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left">Electron monitor unit linearity output</td>
<td valign="top" align="left">Annual</td>
<td valign="top" align="left">&#x00B1;2%</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left">constancy</td>
<td valign="top" align="left"></td>
<td valign="top" align="left">(&#x2265;5 MU)</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left">Dose linearity with dose rate</td>
<td valign="top" align="left">Typical dose rate profile constancy</td>
<td valign="top" align="left">Monthly</td>
<td valign="top" align="left">2%</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left">X-ray output constancy vs. dose rate</td>
<td valign="top" align="left">Annual</td>
<td valign="top" align="left">&#x00B1;2%</td>
<td valign="top" align="left">Change from baseline</td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left">Dose reproducibility with gantry angle</td>
<td valign="top" align="left">X-ray output constancy vs. gantry angle</td>
<td valign="top" align="left">Annual</td>
<td valign="top" align="left">&#x00B1;1%</td>
<td valign="top" align="left">Change from baseline</td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left">Electron output constancy vs. gantry angle</td>
<td valign="top" align="left">Annual</td>
<td valign="top" align="left">&#x00B1;1%</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left">Mechanical test</td>
<td valign="top" align="left">Field light alignment</td>
<td valign="top" align="left">Setting vs. radiation field for two patterns</td>
<td valign="top" align="left">Monthly</td>
<td valign="top" align="left">2 mm</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left">Coincidence of light field and X-ray field</td>
<td valign="top" align="left">Monthly</td>
<td valign="top" align="left">&#x00B1;2 mm</td>
<td valign="top" align="left">All energies</td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left">Cross-hair alignment</td>
<td valign="top" align="left">Cross-hair centering (walkout)</td>
<td valign="top" align="left">Monthly</td>
<td valign="top" align="left">1 mm</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left">Gantry/collimator rotation isocenter</td>
<td valign="top" align="left">Gantry/collimator angle indicators</td>
<td valign="top" align="left">Monthly</td>
<td valign="top" align="left">1&#x00B0;</td>
<td valign="top" align="left">Cardinal angles</td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left">(digital only)</td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left">Collimator rotation isocenter</td>
<td valign="top" align="left">Annual</td>
<td valign="top" align="left">&#x00B1;1 mm</td>
<td valign="top" align="left">Change from baseline</td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left">Gantry rotation isocenter</td>
<td valign="top" align="left">Annual</td>
<td valign="top" align="left">&#x00B1;1 mm</td>
<td valign="top" align="left">Change from baseline</td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left">Optical distance indicator (ODI)</td>
<td valign="top" align="left">Distance indicator (ODI) at isocenter</td>
<td valign="top" align="left">Daily</td>
<td valign="top" align="left">2 mm</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left">verification</td>
<td valign="top" align="left">Distance check device for lasers</td>
<td valign="top" align="left">Monthly</td>
<td valign="top" align="left">1 mm</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left">compared with front pointer</td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left">Collimator rotation readout calibration</td>
<td valign="top" align="left">Gantry/collimator angle indicators</td>
<td valign="top" align="left">Monthly</td>
<td valign="top" align="left">1&#x00B0;</td>
<td valign="top" align="left">Cardinal angles</td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left">(digital only)</td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left">Accessory mount</td>
<td valign="top" align="left">Accessory trays (i.e., port film graticle tray)</td>
<td valign="top" align="left">Monthly</td>
<td valign="top" align="left">2 mm</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left">Latching of wedges, block tray</td>
<td valign="top" align="left">Monthly</td>
<td valign="top" align="left">Functional</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left">Static MLC</td>
<td valign="top" align="left">Leaf position accuracy</td>
<td valign="top" align="left">Qualitative test (i.e., matched segments,</td>
<td valign="top" align="left">Weekly</td>
<td valign="top" align="left">Visual inspection for</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left">aka &#x201C;Picket fence&#x201D;)</td>
<td valign="top" align="left"></td>
<td valign="top" align="left">discernable deviations</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left">such as an increase in</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left">interleaf transmission</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left">Leaf position accuracy</td>
<td valign="top" align="left">Monthly</td>
<td valign="top" align="left">1 mm</td>
<td valign="top" align="left">Four cardinal angles</td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left">Leaf position repeatability</td>
<td valign="top" align="left">Leaf position repeatability</td>
<td valign="top" align="left">Annual</td>
<td valign="top" align="left">&#x00B1;1 mm</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left">Radiation test</td>
<td valign="top" align="left">Collimator spoke shot</td>
<td valign="top" align="left">Leaf spoke shot</td>
<td valign="top" align="left">Annual</td>
<td valign="top" align="left">&#x2264;1.0 mm (radius)</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left">Gantry spoke shot</td>
<td valign="top" align="left">Leaf spoke shot</td>
<td valign="top" align="left">Annual</td>
<td valign="top" align="left">&#x2264;1.0 mm (radius)</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left">Coincidence of light field and X-ray field</td>
<td valign="top" align="left">Coincidence of light field and X-ray field</td>
<td valign="top" align="left">Monthly</td>
<td valign="top" align="left">&#x00B1;2 mm</td>
<td valign="top" align="left">All energies</td>
</tr>
<tr>
<td valign="top" align="left">Dynamic MLC</td>
<td valign="top" align="left">MLC transmission dose rates</td>
<td valign="top" align="left">MLC transmission (average of leaf and</td>
<td valign="top" align="left">Annual</td>
<td valign="top" align="left">&#x00B1;0.5%</td>
<td valign="top" align="left">Change from baseline</td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left">interleaf transmission)</td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left">All energies</td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left">AutoDynalogs for the Millennium MLC</td>
<td valign="top" align="left">-</td>
<td valign="top" align="left">-</td>
<td valign="top" align="left">-</td>
<td valign="top" align="left">Only Varian</td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left">Generate dynalogs for Mark series or</td>
<td valign="top" align="left">-</td>
<td valign="top" align="left">-</td>
<td valign="top" align="left">-</td>
<td valign="top" align="left">Only Varian</td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left">m3 MLC</td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left">RV modeup</td>
<td valign="top" align="left">-</td>
<td valign="top" align="left">-</td>
<td valign="top" align="left">-</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left">Arc dynamic leaf speed test</td>
<td valign="top" align="left">Travel speed</td>
<td valign="top" align="left">Monthly</td>
<td valign="top" align="left">Loss of leaf speed&#x003E;0.5</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left">cm/s</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left">Arc dynamic interlock trip test</td>
<td valign="top" align="left">Arc mode (Expected MU, degrees)</td>
<td valign="top" align="left">Annual</td>
<td valign="top" align="left">&#x00B1;1 mm</td>
<td valign="top" align="left">Change from baseline</td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left">Arc dynamic typical plan test</td>
<td valign="top" align="left">Arc mode (Expected MU, degrees)</td>
<td valign="top" align="left">Annual</td>
<td valign="top" align="left">&#x00B1;1 mm</td>
<td valign="top" align="left">Change from baseline</td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left">Segmental IMRT test (Step and shoot)</td>
<td valign="top" align="left">Segmental IMRT (step and shoot) test</td>
<td valign="top" align="left">Annual</td>
<td valign="top" align="left">RMS maximum of</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left">error&#x003C;0.35 cm</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left">Moving window IMRT test</td>
<td valign="top" align="left">Moving window IMRT</td>
<td valign="top" align="left">Annual</td>
<td valign="top" align="left">RMS maximum of</td>
<td valign="top" align="left">Four cardinal gantry</td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left">error&#x003C;0.35 cm</td>
<td valign="top" align="left">angles</td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left">Moving window IMRT typical plan test</td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left">OBI Safety test</td>
<td valign="top" align="left">Door interlock</td>
<td valign="top" align="left">-</td>
<td valign="top" align="left">-</td>
<td valign="top" align="left">-</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left">Mechanical</td>
<td valign="top" align="left">kV source positioning test</td>
<td valign="top" align="left">Positioning/repositioning</td>
<td valign="top" align="left">Daily</td>
<td valign="top" align="left">&#x2264;2 mm</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left">position accuracy</td>
<td valign="top" align="left">kV Imager positioning test</td>
<td valign="top" align="left">Positioning/repositioning</td>
<td valign="top" align="left">Daily</td>
<td valign="top" align="left">&#x2264;2 mm</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left">verification</td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left">X-ray check</td>
<td valign="top" align="left">Distance measurement demonstration</td>
<td valign="top" align="left">Scaling</td>
<td valign="top" align="left">Monthly</td>
<td valign="top" align="left">Baseline</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left">kV imager panel virtual alignment</td>
<td valign="top" align="left">Imaging and treatment coordinate</td>
<td valign="top" align="left">Daily</td>
<td valign="top" align="left">&#x2264;1 mm</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left">demonstration</td>
<td valign="top" align="left">coincidence</td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left">Optical isocenter demostration</td>
<td valign="top" align="left">Imaging and treatment coordinate</td>
<td valign="top" align="left">Daily</td>
<td valign="top" align="left">&#x2264;2 mm</td>
<td valign="top" align="left">Single gantry angle</td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left">coincidence</td>
<td valign="top" align="left">Monthly</td>
<td valign="top" align="left">&#x2264;2 mm</td>
<td valign="top" align="left">Four cardinal angles</td>
</tr>
<tr>
<td valign="top" align="left">X-ray</td>
<td valign="top" align="left">X-ray measurement-digital fluoroscopy,</td>
<td valign="top" align="left">Imaging dose</td>
<td valign="top" align="left">Annual</td>
<td valign="top" align="left">Baseline</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left">generator</td>
<td valign="top" align="left">pulsed mode (kVp, mA, mS)</td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left">test</td>
<td valign="top" align="left">Digital radiography-dual gain standard</td>
<td valign="top" align="left">Imaging dose</td>
<td valign="top" align="left">Annual</td>
<td valign="top" align="left">Baseline</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left">resolution (kVp, mA, mS)</td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left">HVL using Digital fluoroscopy pulsed</td>
<td valign="top" align="left">Beam quality/energy</td>
<td valign="top" align="left">Annual</td>
<td valign="top" align="left">Baseline</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left">mode</td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left">High contrast resolution</td>
<td valign="top" align="left">Spatial resolution</td>
<td valign="top" align="left">Monthly</td>
<td valign="top" align="left">Baseline</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left">Gray scale linearity</td>
<td valign="top" align="left">Spatial resolution</td>
<td valign="top" align="left">Monthly</td>
<td valign="top" align="left">Baseline</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left">Low contrast sensitivity</td>
<td valign="top" align="left">Contrast</td>
<td valign="top" align="left">Monthly</td>
<td valign="top" align="left">Baseline</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left">Cone beam</td>
<td valign="top" align="left">CT number (Hounsfield unit)</td>
<td valign="top" align="left">HU constancy</td>
<td valign="top" align="left">Monthly</td>
<td valign="top" align="left">Baseline</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left">CT</td>
<td valign="top" align="left">Spatial linearity measurement</td>
<td valign="top" align="left">Geometric distortion</td>
<td valign="top" align="left">Monthly</td>
<td valign="top" align="left">Baseline</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left">(distance)</td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left">Image uniformity measurements</td>
<td valign="top" align="left">Uniformity and noise</td>
<td valign="top" align="left">Monthly</td>
<td valign="top" align="left">Baseline</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left">High resolution</td>
<td valign="top" align="left">Spatial resolution</td>
<td valign="top" align="left">Monthly</td>
<td valign="top" align="left">Baseline</td>
<td valign="top" align="left"></td>
</tr>
<tr>
<td valign="top" align="left"></td>
<td valign="top" align="left">Low contrast resolution</td>
<td valign="top" align="left">Contrast</td>
<td valign="top" align="left">Monthly</td>
<td valign="top" align="left">Baseline</td>
<td valign="top" align="left"></td>
</tr>
</tbody>
</table>
</table-wrap>
</sec>
</back>