Journal List > Ann Dermatol > v.30(1) > 1095390

Lee, Park, Kim, and Kim: Terminal Deoxynucleotidyl Transferase-Mediated Deoxyuridine Triphosphate Nick End Labeling (TUNEL) Assay to Characterize Histopathologic Changes Following Thermal Injury

Abstract

Background

Despite the wide application of lasers and radiofrequency (RF) surgery in dermatology, it is difficult to find studies showing the extent of damage dependent on cell death.

Objective

We evaluated histopathologic changes following in vivo thermal damage generated by CO2 laser, 1,444 nm long-pulsed neodymium:yttrium-aluminum-garnet (LP Nd:YAG) laser and RF emitting electrosurgical unit.

Methods

Thermal damage was induced by the above instruments on ventral skin of rat. Specimens were stained with hematoxylin and eosin, along with a terminal deoxynucleotidyl transferase-mediated deoxyuridine triphosphate nick end labeling (TUNEL) assay, to highlight the degree of irreversible cellular injury.

Results

The volume of vaporization was largest with the CO2 laser. Area of cell death area identified by TUNEL assay, when arranged from widest to narrowest, was 1,444 nm LP Nd:YAG laser, CO2 laser, and RF emitting electrosurgical unit.

Conclusion

This histopathologic evaluation of the acute characterization of injury across devices may be advantageous for attaining better treatment outcomes.

INTRODUCTION

In dermatology, ablative laser and radiofrequency (RF) surgery have been used to replace conventional surgery to remove warts, seborrheic keratoses, various nevi, and benign tumors like hemangiomas, and neurofibromas.
Ablative lasers, the CO2 laser, the 1,444 nm long-pulsed neodymium:yttrium-aluminum-garnet (LP Nd:YAG) laser, and RF emitting electrosurgical units induce a therapeutic effect by heat energy. Infrared light and high-frequency alternating current for laser and RF device, respectively, are converted to heat diffused within the tissues of the operative site. This heat diffusion promotes hemostasis and significantly reduces the bacterial population at the operative site, thus minimizing the risk of intraoperative contamination1.
However, heat diffusion through the operative site results in pain during the procedure and functional damage, as well as potential scarring secondary to permanent tissue transformation. Lateral thermal injury is determined by a number of variables such as hardness and other characteristics of target tissue, as well as wavelength, power, irradiation mode, power density, and frequency of the laser and RF systems utilized.
To enhance therapeutic results, minimize complications and obtain satisfactory cosmetic outcomes, it is important to understand the characteristics of observed histopathologic changes. Although several studies mention histopathologic changes following thermal injury in experiments performed in vivo and in vitro, it is hard to find studies showing extent of damage using cell death as an indicator.
Terminal deoxynucleotidyl transferase-mediated deoxyuridine triphosphate nick end labeling (TUNEL) assay is most widely used to detect DNA damage in situ2. It is not limited to the detection of apoptotic cells: it can also be used to identify necrotic cell death associated with non-apoptotic events such as exposure to toxic compounds and other physical insults3.
Therefore, this preliminary study aims to evaluate histopathological changes following in vivo thermal damage generated by CO2 laser, 1,444 nm LP Nd:YAG laser and RF emitting electrosurgical unit using TUNEL assay.

MATERIALS AND METHODS

Materials

Three six-week-old Sprague-Dawley rats were obtained from Raonbio (Yongin, Korea). All animals were raised in a quiet room maintained on a 12-h light/dark cycle and at 22℃~25℃. They were housed under conventional conditions on wood chip bedding in a Scantainer (Scan-bur BK A/S, Denmark) with two rats per cage with free access to food pellets and water. All experiments were approved by the Korea University College of Medicine Animal Research Policies Committee (KUIACUC-20141117-4).

Methods

1) Thermal injury induced by laser and RF emitting electrosurgical unit

Under anesthesia, we shaved the ventral skin of the rat, divided this area into four regions, and induced thermal injury using an ablative laser and RF emitting electrosurgical unit. The ablative lasers used included the CO2 laser (Ultra CO2 laser; UTI Co., Ltd., Gwangmyeong, Korea) and 1,444 nm LP Nd:YAG laser (AccuSculpt; Lutronic Corporation, Goyang, Korea). The CO2 laser and 1,444 nm LP Nd:YAG laser were irradiated for 2 seconds and parameters of 6.0 watts, continuous wave and 6.0 watts, 40 Hz, 150 mJ, each.
RF emitting electrosurgical unit (Ellman Surgitron FFPF, Hewlett, NY, USA) induced heat damage was also induced for 2 seconds by 6.0 watts (partially rectified/hemo mode) and 31 watts (fully rectified/cut&coagulation mode, a common clinical setting). This mode cuts and coagulates soft tissue by continuous emission of monopolar energy. The RF electrode with a shaft size of 1/16″ (Coagulation/Fulguration F1B; Ellman, Hewlett, NY, USA) was used. The neutral plate was placed on the dependent side of the rat's body.
We aimed to unify the degree of skin ablation between the three devices. The above mentioned energy parameters for each of the devices were selected because they seemed to yield similar depth of skin ablation.

2) Tissue biopsy and staining

All thermally damaged tissues were obtained immediately after irradiation. Incisional biopsies were obtained from the injured sites and specimens were embedded in paraffin. After slides were produced, they were stained with hematoxylin and eosin (H&E) and TUNEL assay was performed to highlight the degree of irreversible cellular injury.

3) TUNEL assay

TUNEL assay was performed with the In Situ Cell Death Detection Kit, Fluorescein (Roche Molecular Biochemicals, Penzberg, Germany) using the manufacturer's instructions. The slides were counterstained with propidium iodide (Vector Laboratories Inc., Burlingame, CA, USA).

4) Light microscopy

The slides stained with H&E were observed at 12.5~200-fold magnification under a light microscope (BX50; Olympus, Tokyo, Japan). Digital photos were taken with a ProgRes microscope camera (Jenoptik, Jena, Germany).

5) Confocal microscopy

The slides analyzed with TUNEL assay were evaluated with confocal microscopy. TUNEL-positive nuclei demonstrated bright green fluorescence using the approximately 488-nm (fluorescein isothiocyanate) fluorescence filter; propidium iodide-positive nuclei demonstrated red fluorescence using the 560-nm filter. Review and photography of slides was carried out at 10~20-fold magnification on Nikon A1-Si confocal laser-scanning microscope (Nikon, Tokyo, Japan).

RESULTS

H&E staining

The tissue damage caused by the CO2 laser showed narrow and deep tissue defects including subcutaneous fat and muscle layer damage by vaporization. Coagulation characterized by basophilic hyalinization and loss of collagen fibrillar texture was evident around the area of the defect (Fig. 1A).
The tissue injured by 1,444 nm LP Nd:YAG laser resulted in a hemispherical tissue defect. Compared to CO2 laser, the vaporized tissue was smaller in volume, but the coagulation zone was more extensive. In addition, changes of the subcutaneous fat layer such as deflation of adipocytes and denaturation of cell membranes were observed (Fig. 1C).
However, the RF emitting electrosurgical unit at 6.0 watts (partially rectified mode), the same power used with the CO2 laser and 1,444 nm LP Nd:YAG laser, didn't result in visible changes on the rat skin. The tissue irradiated at 31 watts (fully rectified mode) on the RF device showed trapezoidal defects that were wider on the epidermal side as well as surrounding coagulation. Compared with CO2 laser and 1,444 nm LP Nd:YAG laser that were irradiated at 6.0 watts, tissue defects were smaller in volume. The depth of the defects didn't surpass the subcutaneous fat layer (Fig. 1E).

TUNEL assay

In the tissue irradiated by CO2 laser, area of cell death surrounding a vaporized area was observed in bright green and yellow fluorescence. Sporadic cell death was also observed in the surrounding tissues (Fig. 1B). Compared to CO2 laser, the RF emitting electrosurgical unit caused cell death restricted only to the vaporized area (Fig. 1F). The tissue irradiated by the 1,444 nm LP Nd:YAG laser showed more extensive thermal injury surrounding the area of the defect than that by the CO2 laser and RF device (Fig. 1D).

DISCUSSION

Several studies compare CO2 laser and RF emitting electrosurgical units during tissue collection and surgery. There is conflicting evidence in studies comparing the lateral thermal injury, level of coagulative necrosis, fibrin deposition, post-operative adhesion, re-epithelialization, post-operative pain and post-operative hemorrhage associated with these treatment modalities14567891011.
Although several pathologic techniques have been developed to evaluate non-viable cells in tissue sections, there is no single method which is widely accepted with sufficient reliability. TUNEL assay remains the most widely used technique to quantify apoptosis with sensitivity between 60% to 90%12. However, the most important drawback of this assay is that it cannot discriminate apoptotic from necrotic cells. Hence we first attempted supravital stain which identifies nicotinamide adenosine dinucleotide phosphate (NADPH). However, this method was too expensive and did not yield consistent result. For these reasons, we decided to use TUNEL assay in order to assess non-viable cells after thermal ablation.
According to the findings of our studies using the TUNEL assay, CO2 laser resulted in more extensive and scattered appearance of cell death area than RF emitting electrosurgical unit. Because both devices are operator-dependent and cannot be standardized, comparison of these methods in clinical settings presents a challenge. Compared to CO2 lasers, RF emitting electrosurgical units are less expensive, do not require eye protection, and are more readily available13.
A 1,444 nm LP Nd:YAG laser has a high affinity to fat and water and has thus been primarily used for laser liposuction procedures14. Now, this laser is being used to treat various conditions such as axillary bromhidrosis1516, skin laxity1718, lipoma19, neurofibroma20, digital mucoid cysts21, and xanthelasmata.
In our study, 1,444 nm LP Nd:YAG laser resulted in smaller vaporized tissue volume and more extensive necrosis and coagulation compared with CO2 laser and RF device. As a pulsed laser, the 1,444 nm LP Nd:YAG laser induces larger accumulated energy, as in doing so it is thought to cause broader damage. In addition, changes in the subcutaneous layer were clearly observed in comparison to the other equipment.
The therapeutic effects from above devices are not only due to thermal effect, but also due to other factors like energy source and specific wavelength. However, in comparison based on the same energy intensity, the volume of vaporization was largest with the CO2 laser. Area of cell death area identified by TUNEL assay, when arranged from widest to narrowest, was 1,444 nm LP Nd:YAG laser, CO2 laser, and RF emitting electrosurgical unit.
TUNEL assay showed histopathologic changes resulting from the various intensities and types of damage that otherwise couldn't be detected by H&E staining. We divided the thermal effects of ablative laser and RF irradiation to tissue into four zones and visualized it as a diagram (Fig. 2). The four zones are composed of vaporization; irreversible tissue damage that appears as cell death and necrosis in the TUNEL assay; reversible tissue damage that appears as denaturation with H&E staining with the absence of necrosis in the TUNEL assay; and, hyperthermia.
However when 1,444 nm LP Nd:YAG laser is applied not to normal skin but to tumors such as neurofibromas, xanthelasmata, angiomas, and pyogenic granulomas, thermal energy is confined within the masses. Consequentially, coagulation and denaturation are predominantly observed while vaporization is minimally observed. Therefore, the characteristics of the treated tissue need to be considered when applying the results of this study clinically.
In addition, the main therapeutic mechanisms of action and strengths and weaknesses of CO2 laser, 1,444 nm LP Nd:YAG laser, and RF device were tabulated (Table 1). When removal of a skin lesion is desired using a modality that replaces surgical treatment, choosing a treatment method in consideration of these points is considered to be helpful to minimize side effects and obtain a good cosmetic result.
There are several limitations to this research. First, in this experiment, rat skin, which has thinner overall thickness than human skin, was used. In particular, the stratum corneum is sparser and the epidermis is thin, so sensitivity to thermal damage is likely to be higher. Currently, there is no experimental model that closely represents human skin apart from the skin of other mammals; development of other models is necessary.
Second, normal skin without pathological lesions was used. When laser and RF devices are applied to human skin, treatment response can vary because treated tissue has various characteristics; for example, acanthotic lesions such as seborrheic keratoses, high-water content tumors like angiomas and neurofibromas, and high-lipid content tumors like xanthelasmata. However, since efficacy and side effects of treatment are influenced significantly by the range of cryogenic and thermal effects to the normal tissue surrounding the lesion, the results of this experiment should be considered for application to practice.
Third, it was difficult to distinguish between apoptotic and necrotic cells using the TUNEL assay. Hereafter, we will consider other feasible stains or serial biopsies on same condition to characterize apoptosis and necrosis.
Fourth, statistical comparison across conditions presented a challenge, as research was not conducted utilizing large sample sizes within each setting.
In conclusion, use of the TUNEL assay made it possible to observe tissue damage at the cellular level that was previously unidentifiable with conventional H&E staining. Thermal damage was variable depending on the device and energy used. Histopathologic analysis of acute cellular damage induced by various devices can be implemented in clinical practice. It could assist in predicting the reaction of human tissues to the planned treatment modality (laser, and RF emitting electrosurgical unit), leading to an optimal treatment outcome. In the future, further studies of cell death with more variable conditions and samples are a requisite.

ACKNOWLEDGMENT

This study was supported by Korea University Grant.

Notes

CONFLICTS OF INTEREST: The authors have nothing to disclose.

References

1. Schoinohoriti OK, Chrysomali E, Tzerbos F, Iatrou I. Comparison of lateral thermal injury and healing of porcine skin incisions performed by CO2-laser, monopolar electrosurgery and radiosurgery: a preliminary study based on histological and immunohistochemical results. Int J Dermatol. 2012; 51:979–986. PMID: 22788819.
crossref
2. Gavrieli Y, Sherman Y, Ben-Sasson SA. Identification of programmed cell death in situ via specific labeling of nuclear DNA fragmentation. J Cell Biol. 1992; 119:493–501. PMID: 1400587.
crossref
3. Ansari B, Coates PJ, Greenstein BD, Hall PA. In situ end-labelling detects DNA strand breaks in apoptosis and other physiological and pathological states. J Pathol. 1993; 170:1–8. PMID: 8326456.
crossref
4. Schemmel M, Haefner HK, Selvaggi SM, Warren JS, Termin CS, Hurd WW. Comparison of the ultrasonic scalpel to CO2 laser and electrosurgery in terms of tissue injury and adhesion formation in a rabbit model. Fertil Steril. 1997; 67:382–386. PMID: 9022618.
crossref
5. Courey MS, Fomin D, Smith T, Huang S, Sanders D, Reinisch L. Histologic and physiologic effects of electrocautery, CO2 laser, and radiofrequency injury in the porcine soft palate. Laryngoscope. 1999; 109:1316–1319. PMID: 10443841.
6. Sinha UK, Gallagher LA. Effects of steel scalpel, ultrasonic scalpel, CO2 laser, and monopolar and bipolar electrosurgery on wound healing in guinea pig oral mucosa. Laryngoscope. 2003; 113:228–236. PMID: 12567074.
crossref
7. Basterra J, Zapater E, Moreno R, Hernández R. Electrosurgical endoscopic cordectomy with microdissection electrodes: a comparative study with CO2 laser. J Laryngol Otol. 2006; 120:661–664. PMID: 16719953.
crossref
8. Silverman EB, Read RW, Boyle CR, Cooper R, Miller WW, McLaughlin RM. Histologic comparison of canine skin biopsies collected using monopolar electrosurgery, CO2 laser, radiowave radiosurgery, skin biopsy punch, and scalpel. Vet Surg. 2007; 36:50–56. PMID: 17214820.
9. Zapater E, Frías S, Pérez A, Basterra J. Comparative study on chronic tissue damage after cordectomies using either CO2 laser or microdissection electrodes. Head Neck. 2009; 31:1477–1481. PMID: 19378326.
10. Stelter K, de la Chaux R, Patscheider M, Olzowy B. Double-blind, randomised, controlled study of post-operative pain in children undergoing radiofrequency tonsillotomy versus laser tonsillotomy. J Laryngol Otol. 2010; 124:880–885. PMID: 20380764.
crossref
11. Schoinohoriti OK, Chrysomali E, Iatrou I, Perrea D. Evaluation of lateral thermal damage and reepithelialization of incisional wounds created by CO2-laser, monopolar electrosurgery, and radiosurgery: a pilot study on porcine oral mucosa. Oral Surg Oral Med Oral Pathol Oral Radiol. 2012; 113:741–747. PMID: 22677020.
crossref
12. Kelly KJ, Sandoval RM, Dunn KW, Molitoris BA, Dagher PC. A novel method to determine specificity and sensitivity of the TUNEL reaction in the quantitation of apoptosis. Am J Physiol Cell Physiol. 2003; 284:C1309–C1318. PMID: 12676658.
13. Taheri A, Mansoori P, Sandoval LF, Feldman SR, Pearce D, Williford PM. Electrosurgery: part II. Technology, applications, and safety of electrosurgical devices. J Am Acad Dermatol. 2014; 70:607.e1–607.e12. quiz 619-620. PMID: 24629362.
14. Tark KC, Jung JE, Song SY. Superior lipolytic effect of the 1,444 nm Nd:YAG laser: comparison with the 1,064 nm Nd:YAG laser. Lasers Surg Med. 2009; 41:721–727. PMID: 20014250.
15. Jung SK, Jang HW, Kim HJ, Lee SG, Lee KG, Kim SY, et al. A prospective, long-term follow-Up study of 1,444 nm Nd:YAG laser: a new modality for treating axillary bromhidrosis. Ann Dermatol. 2014; 26:184–188. PMID: 24882972.
crossref
16. Lee KG, Kim SA, Yi SM, Kim JH, Kim IH. Subdermal coagulation treatment of axillary bromhidrosis by 1,444 nm Nd:YAG laser: a comparison with surgical treatment. Ann Dermatol. 2014; 26:99–102. PMID: 24648694.
crossref
17. Min KH, Kim JH, Park HJ, Chung HS, Heo CY. The skin-tightening effects of 1,444-nm Nd:YAG laser on human skin: an in vivo study. Aesthetic Plast Surg. 2014; 38:585–591. PMID: 24788857.
crossref
18. Lee SH, Roh MR, Jung JY, Jee H, Nam KA, Chung KY. Effect of subdermal 1,444-nm pulsed neodymium-doped yttrium aluminum garnet laser on the nasolabial folds and cheek laxity. Dermatol Surg. 2013; 39:1067–1078. PMID: 23815313.
crossref
19. Lee SH, Jung JY, Roh MR, Chung KY. Treatment of lipomas using a subdermal 1,444-nm micropulsed neodymiumdoped yttrium aluminum garnet laser. Dermatol Surg. 2011; 37:1375–1376. PMID: 22988996.
20. Kim HJ, Lee KG, Yi SM, Kim JH, Kim IH. Successful treatment of multiple cutaneous neurofibromas using a combination of shave excision and laser photothermocoagulation with a 1,444-nm neodymium-doped yttrium aluminum garnet laser. Dermatol Surg. 2012; 38:960–963. PMID: 22329357.
crossref
21. Kim JH, Park JH, Jee H, Oh SH. Successful treatment of recurrent digital mucoid cysts using a 1,444-nm neodymium-doped yttrium aluminum garnet laser. Dermatol Surg. 2011; 37:1528–1530. PMID: 21707830.
crossref
Fig. 1

Corresponding hematoxylin and eosin (H&E; A, C, E, ×20) and terminal deoxynucleotidyl transferase-mediated deoxyuridine triphosphate nick-end labeling (TUNEL)-stained (B, D, F, ×20) histological skin sections following treatment with the CO2 laser (A, B), 1,444 nm long-pulsed neodymium:yttrium-aluminum-garnet (C, D), and radiofrequency emitting electrosurgical unit (E, F). Note the bright yellow/green cells in the TUNEL-stained sections highlighting the necrotic cells within the treated section.

ad-30-41-g001
Fig. 2

Thermal effects of ablative laser and radiofrequency irradiation. (A) CO2 laser and radiofrequency emitting electrosurgical unit. (B) 1,444 nm long-pulsed neodymium:yttrium-aluminum-garnet (LP Nd:YAG). 1,444 nm LP Nd:YAG laser resulted in smaller vaporized tissue volume and more extensive necrosis and coagulation compared with CO2 laser and radiofrequency device.

ad-30-41-g002
Table 1

Comparison of the characteristics and advantages/disadvantages of the three treatments modalities

ad-30-41-i001
Characteristic CO2 laser 1,444 nm LP Nd:YAG laser Radiofrequency emitting electrosurgical unit
Mode of action Vaporization Coagulative necrosis, denaturation Thermal necrosis
Selectivity for the tissue Non-specific tissue ablation Superior lipolytic effect Non-specific tissue ablation
Need for local anesthesia + + +
Bleeding during treatment ++ +
Procedure time Short Moderate Short
Need for suction + +
Need for eye protection + +
Cost of equipment Expensive Expensive Less expensive
Complication Hyper or hypopigmentation, hypertrophic scar Hyper or hypopigmentation, hypertrophic scar Hyper or hypopigmentation, hypertrophic scar

LP Nd:YAG: long-pulsed neodymium:yttrium-aluminum-garnet.

TOOLS
Similar articles