Shin, Choi, Kim, and Jeong: Isocenter Check and QA of Tactic Radiosurgery Devices Using EPID

Abstract

The Winston Lutz test, which checks the accuracy of the isocenter for stereotactic radiosurgery (SRS), was performed with the commercial electronic portal imaging device (EPID). The usual Winston Lutz test with film was also performed for comparison with the test with EPID. The maximum difference in isocenter between the two methods was 0.32 mm. The Winston Lutz test using EPID is practical as it can reduce time and avoid human errors compared to the test with film.

References

1. Lutz W, Winston KR, Maleki N. A system for stereotactic radiosurgery with a linear accelerator. Int J Radiat Oncol Biol Phys. 1988; 14(2):373–81.
crossref
2. Podgorsak EB, Pike GB, Olivier A, Pla M, Souhami L. Radiosurgery with high energy photon beams: a comparison among techniques. Int J Radiat Oncol Biol Phys. 1989; 16(3):857–65.
crossref
3. Podgorsak EB, Pike GB, Pla M, Olivier A, Souhami L. Radiosurgery with photon beams: physical aspects and adequacy of linear accelerators. Radiother Oncol. 1990; 17(4):349–58.
crossref
4. Ramaseshan R and Heydarian M. Comprehensive quality assurance for stereotactic radiosurgery treatments. Phys Med Biol. 2003; 48(14):N199–N205.
5. Verellen D, Linthout N, Bel A, et al. Assessment of the uncertainties in dose delivery of a commercial system for linac-based stereotactic radiosurgery. Int J Radiat Oncol Biol Phys. 1999; 44(2):421–33.
crossref
6. Serago ChF. Lewin AA, Houdek PV, et al. Radiosurgery target point alignment errors detected with portal film verification. Int J Radiat Oncol Biol Phys. 1992; 24(4):777–80.
7. Leksell L. The stereotaxic method and radiosurgery of the brain. Acta Chir Scand. 1951; 102(4):316–19.
8. Pejman Rowshanfarzad, Mahsheed Sabet, Daryi J. OConnor, Peter B. Greer, et, al: Isocenter verification for linac-based stereotactic radiation therapy: review of principles and techniques. J of Appl. Clini. Med. Phy. 2011; 12(4):
9. E. Shaw, C. Scott, L. Souhami, et al: Single dose radiosurgical treatment of recurrent previously irradiated primary brain tumors and brain metastases: Final report of RTOG protocol 90–05 Int J Radiat Oncol Biol Phys, 47 (. 2000. ), pp.291–298.
10. L. Walton, A. Hampshire, A. Roper, et al: Development of a relocatable frame technique for gamma knife radiosurgery: Technical note J Neurosurg, 93 (. 2000. ), pp.198–202.
11. H. Shirato, T. Sakamoto, N. Takeuchi, et al: Fractionated stereotactic radiotherapy for vestibular schwannoma (vs): Comparison between cystic-type and solid-type vs, Int J Radiat Oncol Biol Phys, 48 (. 2000. ), pp.1395–1401.
12. J.A. Williams: Fractionated stereotactic radiotherapy for acoustic neuromas, Int J Radiat Oncol Biol Phys, 54 (. 2002. ), pp.500–504.
13. Zheng C, Zhiheng W, Jinli M, et al. Six degree-of-freedom image guidance for frameless intracranial stereotactic radiosurgery with kilovoltage cone-beam CT, J Nucl Med Radiat Ther. 2010; 1:101.

Fig. 1.
Clinical linear accelerator (RapidArc Varian Co.).
pmp-25-281f1.tif
Fig. 2.
Image of micro-multileaf collimator (m3, BrainLab).
pmp-25-281f2.tif
Fig. 3.
Electronic portal imaging device (EPID, Varian Co.).
pmp-25-281f3.tif
Fig. 4.
Imge of Winston Lutz Test Tool Sets.
pmp-25-281f4.tif
Fig. 5.
The setup for a Winston Lutz Test ball at the isocenter.
pmp-25-281f5.tif
Fig. 6.
The image of film setup for Winston Lutz Test at gantry 0o and Table 0o.
pmp-25-281f6.tif
Fig. 7.
The image of film setup for Winston Lutz Test at gantry 90o and Table 0o.
pmp-25-281f7.tif
Fig. 8.
The image of EPID setup for Winston Lutz Test at gantry 0o and Table 0o.
pmp-25-281f8.tif
Fig. 9.
The image of EPID setup for Winston Lutz Test at gantry 180o and Table 0o.
pmp-25-281f9.tif
Fig. 10.
The results of Winston Lutz Test for clinical linear accelerator.
pmp-25-281f10.tif
Fig. 11.
Film of a passed Winstion-Lutz test m3 for radiation field and isocenter for FSRT.
pmp-25-281f11.tif
Fig. 12.
EPID results of a passed Winstion-Lutz test m3 for radiation field and isocenter.
pmp-25-281f12.tif
Table 1.
The results of film for winston lutz test at various gantry and couch angles.
Gantry Table Δx Δy X G=0 G=90 G=180 
T=0 T=90 T=270 T=0 T=0 T=90 T=270
  (mm) X (mm) Y (mm) X (mm) Y (mm) X (mm) Y (mm) X (mm) Y (mm) X (mm) Y (mm) X (mm) Y (mm)x (mm) Y
  0.10 −0.60 −0.10 0.50 0.80 0.00 0.01 − −0.01 0.00 0.60 1.10 1.10 − −0.03 0.30
  0.30 −0.20 1.10 −0.01 0.01 −0.40 0.80 − −0.10 0.30 0.15 1.10 1.10 − −0.30 0.50
  −0.01 −0.80 −0.30 0.85 1.10 −0.01 0.30 − −0.25 0.60 0.15 1.10 1.10 − −0.40 0.25
  0.01 −0.30 −0.20 0.90 1.10 0.01 0.15 − −0.01 0.20 0.00 1.10 1.10 − −0.20 0.60
  −0.01 0.10 −0.20 0.80 0.80 0.00 0.15 0.01 0.70 0.05 0.90 1.10 − −0.15 0.40
Mean 0.08 −0.36  0.06 0.61  0.76 −0.08 0.28 − −0.07 0.36 0.19 1.06 1.10 − −0.22 0.41
St.Dev. 0.13 0.35 0.59 0.38 0.45 0.18 0.31 0.11 0.29 0.24 0.09 0.00 0.14 0.14
Min −0.01 −0.80 −0.30 −0.01 0.01 −0.40 0.01 − −0.25 0.00 0.00 0.90 1.10 − −0.40 0.25
Max  0.30 0.10 1.10 0.90  1.10 0.01 0.80 0.01 0.70 0.60 1.10 1.10  − −0.03 0.60
Table 2.
The results of EPID images for winston lutz test at various gantry and couch angles.
Gantry Table Δx Δy X G=0 G=90 G=180 
T=0 T=90 T=270 T=0 T=0 T=90 T=270
  X (mm) Y (mm) X (mm) Y (mm) X (mm) Y (mm) X (mm) Y (mm) X (mm) Y (mm) X (mm) Y (mm) X (mm) Y (mm)
  0.01 −0.05 0.01 −0.90 0.50 −0.10 −0.15 −0.10 −0.40 0.00 0.50 0.50 −0.15 0.01
  −0.01 −0.90 −0.20 0.01 0.20 −0.60 0.00 −0.15 0.00 −0.01 0.80 0.85 −0.01 0.25
  0.00 −0.80 −0.30 0.85 1.00 −0.01 0.30 −0.10 0.30 0.10 1.00 1.00 −0.20 0.30
  0.01 −0.30 −0.05 0.70 0.70 0.30 0.20 −0.15 0.40 0.40 1.10 0.90 −0.30 0.40
  −0.01 −0.40 −0.10 0.80 0.85 0.05 0.10 0.05 0.10 0.30 1.00 1.00    −0.15 0.50
Mean  0.00 −0.49 −0.13 0.29 0.65 −0.07 0.09 −0.09 0.08 0.16 0.88 0.85 −0.16 0.29
St. Dev.  0.01 0.35 0.12 0.75 0.31 0.33 0.17 0.08 0.31 0.18 0.24 0.21 0.10 0.18
Min −0.01 −0.90 −0.30 −0.90 0.20 −0.60 −0.15 −0.15 −0.40 −0.01 0.50 0.50 −0.30 0.01
Max 0.01 −0.05 0.01 0.85 1.00 0.30 0.30 0.05 0.40 0.40 1.10 1.00 −0.01 0.50
Table 3.
Difference of results in film and EPID for winston lutz test at varius gantry and couch angles.
Gantry Table Δx Δy X G=0 G=90 G=180 
T=0 T=90 T=270 T=0 T=0 T=90 T=270
  X (mm) Y (mm) X (mm) Y (mm) X (mm) Y (mm) X (mm) Y (mm) X (mm) Y (mm) X (mm) Y (mm) X (mm) Y (mm
Film  0.08 −0.36     0.06 0.61   0.76 −0.08 0.28 −0.07 0.36 0.19 1.06 1.10 −0.22 0.41
EPID 0.00 −0.49 −0.13 0.29 0.65 −0.07 0.09 −0.09 0.08 0.16 0.88 0.85 −0.16 0.29
Diff.  0.08 0.13     0.19 0.32   0.11 −0.01 0.19 0.02 0.28 0.03 0.18 0.25 −0.05 0.12